Recall Record

Pleasanton, CA, United States

Calyxo, Inc. recalls

1 recall recorded since 2018, most recently on 19 February 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
19 February 2025
Reason
Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1396-2025