Recall Record

Pleasanton, CA, United States

CAIRE DIAGNOSTICS INC recalls

1 recall recorded since 2018, most recently on 26 February 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
26 February 2021
Terminated
24 March 2023
Reason
Erroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease. Two issues affecting the FeNO readings: 1) Drift in the calibration gases and 2) Software error. Each of these issues can cause FeNO scores/results to be erroneously high.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of CA, CO, FL, TX, and WI. In the countries of Belgium, Canada, Croatia/Hrvatska, France, Germany, Greece, Hong Kong, India, Philippines, Spain, Spain and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1284-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.