Breas Medical, Inc. recalls

5 recalls recorded since 2018, most recently on 25 July 2024. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)3
Terminated or completed2
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
25 July 2024
Reason
There is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in specific conditions.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-2690-2024

Ongoing Class II Medical devices

Recall started
13 October 2023
Reason
Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
Distribution
Nationwide Foreign: Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-0405-2024

Ongoing Class II Medical devices

Recall started
2 June 2022
Reason
A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1382-2022

Terminated Class II Medical devices

Recall started
21 March 2022
Terminated
13 December 2024
Reason
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0943-2022

Terminated Class II Medical devices

Recall started
28 January 2020
Terminated
15 September 2020
Reason
The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, CA, FL, GA, IA, IL, KS, MA, MI, MO, NJ, NY, OH, PR, SC, TX, VA, WI and countries distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1352-2020