Recall Record

Hazelwood, MO, United States

Biomerieux Inc recalls

39 recalls recorded since 2018, most recently on 24 July 2026. 12 are listed as ongoing. Status and classification are as published by the FDA.

Recalls39
Ongoing (FDA)12
Terminated or completed27
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
24 July 2026
Reason
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Distribution
International distribution to the countries of Austria, Germany, Luxembourg.
Initiated
Voluntary: Firm initiated
Recall number
Z-3231-2026

Ongoing Class II Medical devices

Recall started
4 June 2026
Reason
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
Distribution
International distribution to the countries of Belgium and the Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-2715-2026

Ongoing Class II Medical devices

Recall started
5 January 2026
Reason
Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1236-2026

Ongoing Class II Medical devices

Recall started
26 June 2025
Reason
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Distribution
International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Polan
Initiated
Voluntary: Firm initiated
Recall number
Z-2671-2025

Ongoing Class I Medical devices

Recall started
14 March 2024
Reason
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1464-2024

Ongoing Class II Medical devices

Recall started
6 February 2023
Reason
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.
Distribution
US Nationwide distribution to the states of Alabama, California, North Carolina, and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-1201-2023

Ongoing Class II Medical devices

Recall started
9 November 2022
Reason
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0873-2023, Z-0798-2023, Z-0864-2023, Z-0839-2023, Z-0837-2023, Z-0817-2023, Z-0829-2023, Z-0856-2023, Z-0806-2023, Z-0827-2023, Z-0851-2023, Z-0863-2023 and 82 more

Ongoing Class II Medical devices

Recall started
20 July 2022
Reason
There is a potential of no result leading to possible delayed result due to contamination of extraction reagents.
Distribution
Distribution was made to AR, AZ, CA, CT, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NJ, NY, OH, OR, PA, TN, TX, and VA. There was no government/military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1564-2022

Ongoing Class II Medical devices

Recall started
13 July 2022
Reason
There are 7 reported software anomalies that may affect use of the device.
Distribution
International distribution to the countries of Estonia, Bahrain, Pakistan, UK, Singapore, Saudi Arabia, Thailand, and Vietnam.
Initiated
Voluntary: Firm initiated
Recall number
Z-1640-2022

Ongoing Class II Medical devices

Recall started
29 March 2022
Reason
Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.
Distribution
U.S. distribution of the FSCA to NC. OUS distribution of the FSCA to subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Polan
Initiated
Voluntary: Firm initiated
Recall number
Z-1059-2022

Terminated Class II Medical devices

Recall started
9 February 2022
Terminated
1 April 2026
Reason
Invalid calibrations (out of range high) while using the product.
Distribution
Distribution was made to CA, FL, GA, IL, IN, LA, MA, MD, ME, MO, NJ, NY, OH, PA, TX, VA, WI, and WV. There was government distribution but no military or foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0742-2022

Terminated Class II Medical devices

Recall started
20 December 2021
Terminated
25 September 2023
Reason
Risk of misidentification or delayed results with the absence of identification due to thermoforming defect.
Distribution
US (Domestic) Distribution to states of: AZ, IA, KY, and NY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0525-2022

Terminated Class II Medical devices

Recall started
9 November 2021
Terminated
10 February 2025
Reason
There is an error in the new version of the package insert. The Reading Table lists the incorrect color for a negative result for the CIT test.
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MO, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, and WI. There was government distribution but no military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0422-2022

Terminated Class II Medical devices

Recall started
30 September 2021
Terminated
18 January 2024
Reason
There is potential to provide an incorrect organism identification.
Distribution
Distribution in WI, NY, and SD
Initiated
Voluntary: Firm initiated
Recall number
Z-0243-2022

Ongoing Class II Medical devices

Recall started
22 September 2021
Reason
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0432-2022, Z-0434-2022, Z-0438-2022, Z-0431-2022, Z-0435-2022, Z-0437-2022, Z-0424-2022, Z-0433-2022, Z-0430-2022, Z-0426-2022, Z-0427-2022, Z-0436-2022 and 4 more

Terminated Class II Medical devices

Recall started
2 September 2021
Terminated
9 March 2026
Reason
Under certain conditions, there is a risk for a false negative result.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Australia, Belgium, Canada, Switzerland, Chile, Columbia, Czech Republic, Denmark, Germany, Spain, Finland, France, United Kingdom, Guadeloupe, Hong Kong, Canary Islands, Ireland, India, Italy, Japan, South Korea, The
Initiated
Voluntary: Firm initiated
Recall number
Z-0084-2022

Terminated Class II Medical devices

Recall started
24 June 2021
Terminated
10 February 2025
Reason
Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii. The countries of Austria, Belgium, Croatia, Czech Republic, France, Germa
Initiated
Voluntary: Firm initiated
Recall number
Z-2226-2021

Terminated Class II Medical devices

Recall started
22 June 2021
Terminated
6 February 2023
Reason
Software anomaly - Under certain conditions, unwanted alterations to results could be applied when using the BCI Connect scripting feature.
Distribution
Distribution in the U.S. was nationwide. There was also military and government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2225-2021

Terminated Class II Medical devices

Recall started
16 March 2021
Terminated
6 August 2024
Reason
bioMerieux received complaints about calibration issue observed on several lots of VIDAS¿ CMV IgM (ref 30205 ; 30205-01).
Distribution
Distributed nationwide to NY, CA, IL, WV, NJ, NY, TX, IN, PA, WI, VA, GA, CO, MI, FL, OH, MA, LA, NV, MO, CT and internationally to Argentina, Australia, Belgium, Brazil, Switzerland, Chile, Colombia, Czech Republic, Germany, Egypt, Spain, France, United Kingdom, Greece, Hong Kong, Hungary, India,
Initiated
Voluntary: Firm initiated
Recall number
Z-1408-2021

Terminated Class II Medical devices

Recall started
28 January 2021
Terminated
12 January 2023
Reason
bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK¿ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK¿ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).
Distribution
Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, L
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1308-2021, Z-1307-2021

Ongoing Class II Medical devices

Recall started
11 March 2020
Reason
There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card (ref. 421857, lot 8021232403). .
Distribution
No distribution within the United States. Foreign distribution locations: Netherlands, Luxembourg, Belgium
Initiated
Voluntary: Firm initiated
Recall number
Z-1742-2020

Terminated Class II Medical devices

Recall started
5 February 2020
Terminated
28 October 2022
Reason
The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card.
Distribution
International distribution in the countries of Germany and the Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-1534-2020

Terminated Class II Medical devices

Recall started
24 January 2020
Terminated
11 January 2023
Reason
The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.
Distribution
Worldwide distribution. US nationwide including AU, BE, CH, DE, ES, FR, IT, HK/CN; MX, NL, RU, SE/FI, ZA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1793-2020

Terminated Class II Medical devices

Recall started
10 December 2019
Terminated
16 February 2021
Reason
Two batches of VIDAS Varicella-Zoster IgG (VZG ref 30217) may yield an invalid calibration which may delay the patient's sample result or a valid calibration may lead to a false positive patient result.
Distribution
Distributed nationwide and 9 units to Biomerieux Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1107-2020

Terminated Class II Medical devices

Recall started
22 October 2019
Terminated
6 August 2024
Reason
Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall number
Z-0928-2020

Terminated Class II Medical devices

Recall started
15 March 2019
Terminated
17 August 2021
Reason
Unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits.
Distribution
Internationally to: United Kingdom, Latvia, Poland, Ukraine, Germany, and Scotland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1137-2019

Terminated Class II Medical devices

Recall started
13 December 2018
Terminated
16 August 2021
Reason
False Positive ESBL Phenotype
Distribution
UK and Ireland
Initiated
Voluntary: Firm initiated
Recall number
Z-0847-2019

Terminated Class II Medical devices

Recall started
13 November 2018
Terminated
7 July 2021
Reason
The VITEK¿ 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) results when using the VITEK¿ 2 FLEXPREP" feature.
Distribution
Nationwide distribution to CO, IL, MO, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0639-2019

Terminated Class II Medical devices

Recall started
22 October 2018
Terminated
4 November 2019
Reason
False resistant results for Streptococcus anginosus and Streptococcus constellatus strains were reported.
Distribution
Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY including Puerto Rico and to countries of:Algeria,
Initiated
Voluntary: Firm initiated
Recall number
Z-0618-2019

Terminated Class II Medical devices

Recall started
8 October 2018
Terminated
21 May 2021
Reason
Some eluates become colored because of remaining hemoglobin in the eluate.
Distribution
CA, FL, GA, MD, MI, MN, NC, NJ, NM, NY, OH, OR, RI, SC, TN, TX, VA, WA, and WI Austria, Australia, Belgium, Switzerland, Colombia, Germany, Spain, France, UK, India, Italy, Japan, Netherlands, Portugal, Sweden, Singapore, South Africa, Israel, Dutch Antilles, Angola, Estonia, and Bangladesh
Initiated
Voluntary: Firm initiated
Recall number
Z-0443-2019

Terminated Class II Medical devices

Recall started
17 September 2018
Terminated
18 May 2020
Reason
Data loggers showed that one shipment to the United States experienced low temperatures that may impact performance of this product.
Distribution
Nationwide distribution to MD, TX, MA, GA, OH, OR, CA, and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0218-2019

Terminated Class II Medical devices

Recall started
13 August 2018
Terminated
2 November 2020
Reason
Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).
Distribution
Worldwide Distribution US Nationwide and the countries of: Turkey, Austria, Switzerland, Czech Republic, Germany, Netherlands,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3180-2018, Z-3179-2018

Terminated Class II Medical devices

Recall started
27 June 2018
Terminated
21 July 2020
Reason
The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KS, LA, MI, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VT, WV. Distributed internationally to Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Germany, Spain, France, Uni
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0229-2019, Z-0230-2019, Z-0228-2019

Terminated Class II Medical devices

Recall started
9 April 2018
Terminated
2 December 2025
Reason
Invalid calibration with low calibrator S1 while using the product.
Distribution
Distribution was made to the recalling firm's distribution center in KY. There was no military/government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2331-2018

Terminated Class II Medical devices

Recall started
23 March 2018
Terminated
23 June 2020
Reason
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Distribution
International distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2167-2018, Z-2166-2018, Z-2169-2018, Z-2164-2018, Z-2165-2018, Z-2168-2018, Z-2172-2018, Z-2170-2018, Z-2171-2018

Terminated Class II Medical devices

Recall started
1 March 2018
Terminated
9 December 2025
Reason
Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.
Distribution
Worldwide Distribution - US Nationwide in the state of Arizona and North Carolina. Distributed internationally to Algeria, Algeria, Angola, Argentina, Bahrain, Bangladesh, Belarus, Bosnia-Herz., Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chad, Colombia, Congo, Costa Rica, Cote d'Ivoire,
Initiated
Voluntary: Firm initiated
Recall number
Z-2588-2018

Terminated Class I Medical devices

Recall started
26 January 2018
Terminated
20 May 2019
Reason
Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Bermuda, Brazil, Canada, China, Curacao, Costa Rica, Dominican Republic, Ecuador, France, Guatemala, Indonesia, Israel, Hong Kong, South Korea, South Africa, Mexico, Myanmar, Nicaragua, Peru, Philippines, Paraguay, Singapore, Thailand, El
Initiated
Voluntary: Firm initiated
Recall number
Z-2817-2018

Terminated Class II Medical devices

Recall started
23 January 2018
Terminated
13 July 2021
Reason
Missing desiccant.
Distribution
US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT. VA. VT. WA. WI, and WV
Initiated
Voluntary: Firm initiated
Recall number
Z-2874-2018

Terminated Class II Medical devices

Recall started
4 January 2018
Terminated
14 April 2022
Reason
Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS¿ and mini VIDAS¿ system due to a decrease of pipetted volume.
Distribution
Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY. International distribution to Algeria, Angola, Angu
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2429-2018, Z-2428-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.