Biomeme, Inc. recalls

3 recalls recorded since 2018, most recently on 20 October 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)0
Terminated or completed3
Class I (FDA)0

Terminated Class II Medical devices

Recall started
20 October 2021
Terminated
5 March 2024
Reason
The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).
Distribution
Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-0452-2022

Terminated Class III Medical devices

Recall started
5 January 2021
Terminated
11 June 2021
Reason
Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.
Distribution
Domestic: FL, TX, IN, KY, OH, NY, CA, PA, MD. Foreign: Canada, Zimbabwe.
Initiated
Voluntary: Firm initiated
Recall number
Z-0940-2021

Terminated Class II Medical devices

Recall started
23 October 2020
Terminated
25 January 2021
Reason
The firm has become aware of nine reports by a single customer that the software made false positive calls. The investigation determined that the root cause is the higher than expected amber-to-red co-excitation in the assay.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, IN, NY, OH, PA, and the countries of Canada, Singapore, Zimbabwe, and Dominican Republic.
Initiated
Voluntary: Firm initiated
Recall number
Z-0513-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.