BioMedical Equipment Service Co (BMES) recalls

1 recall recorded since 2018, most recently on 25 February 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)1

Terminated Class I Medical devices

Recall started
25 February 2021
Terminated
22 April 2024
Reason
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Distribution
The affected products were distributed to the following US states: CA, IL, IN, KY, LA, NJ, NV, OH, and PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2204-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.