Pacheco, CA, United States

Biocare Medical, LLC recalls

5 recalls recorded since 2018, most recently on 25 September 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)1
Terminated or completed4
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
25 September 2024
Reason
Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.
Distribution
US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0315-2025

Terminated Class II Medical devices

Recall started
25 February 2021
Terminated
31 January 2023
Reason
Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.
Distribution
US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands
Initiated
Voluntary: Firm initiated
Recall number
Z-1333-2021

Terminated Class II Medical devices

Recall started
10 October 2019
Terminated
2 December 2020
Reason
Firm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting. Positive control would observe no staining. The clinical impact may result in the consumption of patient tissue sample and a delay in diagnosis.
Distribution
US: TX, MA, TN, MO, IN
Initiated
Voluntary: Firm initiated
Recall number
Z-2291-2020

Terminated Class II Medical devices

Recall started
24 April 2018
Terminated
14 July 2020
Reason
Chromogen failed to develop due to a stability issue caused by environmental factors with a raw material in the Vulcan Fast Red Buffer.
Distribution
CA, VT, KY, LA, OR, RI, GA, IL, MA, TX, KS, FL, Japan, and Germany
Initiated
Voluntary: Firm initiated
Recall number
Z-2360-2018

Terminated Class II Medical devices

Recall started
15 March 2018
Terminated
22 April 2020
Reason
Immunohistochemistry device kits found to display incorrect staining pattern due to mislabeling of two components.
Distribution
Worldwide Distribution - Unites States Nationwide in the states of NY AND FL and the countries of: Morocco and Italy
Initiated
Voluntary: Firm initiated
Recall number
Z-3198-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.