Bio-Rad Laboratories, Inc. recalls

14 recalls recorded since 2018, most recently on 30 May 2025. 11 are listed as ongoing. Status and classification are as published by the FDA.

Recalls14
Ongoing (FDA)11
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
30 May 2025
Reason
Due to a risk of false positive results that could lead to unnecessary medical treatment.
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, NC and NV.
Initiated
Voluntary: Firm initiated
Recall number
Z-2515-2025

Ongoing Class II Medical devices

Recall started
15 February 2024
Reason
Due to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.
Distribution
US Nationwide distribution in the states of PA, SD, TX, FL, MN.
Initiated
Voluntary: Firm initiated
Recall number
Z-1358-2024

Ongoing Class II Medical devices

Recall started
28 February 2023
Reason
APLS IgM reagent kits were packaged with the incorrect conjugate, which could lead to an increase in false-positive and false-negative results.
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, FL, HI, IL, IN, KS, KY, LA, MI, MO, NC, ND, NE, NJ, NM, NY, OH, OK, PR,SC, TN, TX, UT, VT.
Initiated
Voluntary: Firm initiated
Recall number
Z-1572-2023

Ongoing Class II Medical devices

Recall started
21 February 2023
Reason
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1  98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1  98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1  98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, Germany, Spain, France, United Kingdom, Israel, Italy, Norway, Sweden.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1378-2023, Z-1377-2023

Ongoing Class II Medical devices

Recall started
19 October 2022
Reason
Due to an unusual increase in the positivity rate with human IgM antibodies to cytomegalovirus (CMV) Enzyme Immunoassay (EIA).
Distribution
U.S.: CA, CO, FL, GA, HI, IA, MA, MN, NC, NH, NJ, PA, TN, and TX O.U.S.: None
Initiated
Voluntary: Firm initiated
Recall number
Z-0342-2023

Ongoing Class II Medical devices

Recall started
14 April 2022
Reason
Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ, CA, FL, HI, IA, IL, MI, MN, NE, NY, TN, TX, and WA. The countries of Czech Republic, France, Germany, and Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-1096-2022

Ongoing Class II Medical devices

Recall started
7 April 2022
Reason
The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CD¿ROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall.
Distribution
US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR OUS: Canada, Spain, France, Italy, India, Vietnam, Philippines
Initiated
Voluntary: Firm initiated
Recall number
Z-1208-2022

Ongoing Class II Medical devices

Recall started
8 February 2022
Reason
Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet stability specifications
Distribution
Worldwide distribution U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of France, Germany, Italy, Spain, Sa
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0806-2022, Z-0807-2022

Ongoing Class II Medical devices

Recall started
17 September 2021
Reason
Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.
Distribution
Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0236-2022, Z-0235-2022

Ongoing Class II Medical devices

Recall started
21 April 2020
Reason
The increased level of RPR reactivity in the BioPlex 2200 Syphilis Total & RPR kit material number 12000650.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL,GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TX, UT, VA, WA, WI, and WV. The countries of Canada (CA), Switzerland (CH), Fr
Initiated
Voluntary: Firm initiated
Recall number
Z-2982-2020

Terminated Class II Medical devices

Recall started
1 November 2019
Terminated
27 September 2023
Reason
A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.
Distribution
Worldwide distribution - U.S Nationwide distribution including in the states of CA, FL, NJ, IL, NE, TN, IN, TX, LA, WA, MI, MA, MN, OH, HI, MO, PA, CT, UT, SD, NY, AL, PR, OR, MD, KY, NC, AZ, SC, VA, WI, GA, IA, RI, AK, OK, KS, ND, MS, ME, NM, AR and the countries of ES, AE, DE, FR, KR, CN, TW, JP,
Initiated
Voluntary: Firm initiated
Recall number
Z-0082-2022

Ongoing Class II Medical devices

Recall started
10 January 2019
Reason
User-induced circumstances can contribute to the EVOLIS Microplate Processor System not handling repeat testing requests/results properly and may contribute to transmitting erroneous results to the customer's Laboratory Information System (LIS).
Distribution
Distribution was nationwide to medical facilities, including to Puerto Rico. There was also government/military distribution. There was no foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1906-2019

Terminated Class II Medical devices

Recall started
8 June 2018
Terminated
30 October 2020
Reason
Abnormal coloration of agar plates and the growth of non-characteristic MRSA colonies.
Distribution
US Nationwide Distribution to: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1742-2019, Z-1743-2019

Terminated Class II Medical devices

Recall started
2 April 2018
Reason
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.
Distribution
US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-2541-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.