Irvine, CA, United States
Beta Bionics, Inc. recalls
3 recalls recorded since 2018, most recently on 12 March 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.
Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 12 March 2026
- Reason
- insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
- Distribution
- US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1996-2026
- iLet Bionic Pancreas, REF: BB1001
Ongoing Class II Medical devices
- Recall started
- 23 September 2025
- Reason
- The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
- Distribution
- US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2513-2026
- iLet Bionic Pancreas, REF: BB1001
Ongoing Class II Medical devices
- Recall started
- 13 August 2025
- Reason
- it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1809-2026
- iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software