Beckman Coulter Inc. recalls
128 recalls recorded since 2018, most recently on 2 September 2026. 81 are listed as ongoing. Status and classification are as published by the FDA.
Recalls128
Ongoing (FDA)81
Terminated or completed47
Class I (FDA)2
Ongoing Class II Medical devices
- Recall started
- 2 September 2026
- Reason
- Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0083-2027, Z-0084-2027
- DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252
- DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
Ongoing Class II Medical devices
- Recall started
- 12 August 2026
- Reason
- The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
- Distribution
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the c
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0160-2027, Z-0161-2027
- DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
- IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
Ongoing Class II Medical devices
- Recall started
- 15 July 2026
- Reason
- a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
- Distribution
- Worldwide distribution - US Nationwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2974-2026, Z-2973-2026
- Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
- Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Mod
Ongoing Class II Medical devices
- Recall started
- 26 June 2026
- Reason
- Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
- Distribution
- US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2889-2026, Z-2888-2026
- REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan Walk
- REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284
Ongoing Class II Medical devices
- Recall started
- 16 March 2026
- Reason
- Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bangladesh, Bolivia (Plurinational State of), Bosnia and Herzegovina, Brazil, Canada, China, C¿te d'Ivoire, Croatia, Czechia, Denmark, Egypt, El Salvador, Ethiopia, Georgia, Germany, Greece, Hong Kong, India, Indonesia,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1922-2026
- Access Total T4 Calibrator, Catalog No. 33805
Ongoing Class II Medical devices
- Recall started
- 23 December 2025
- Reason
- Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
- Distribution
- Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1235-2026
- CellMek SPS Sample Preparation System, REF: C44603, with software
Ongoing Class II Medical devices
- Recall started
- 26 November 2025
- Reason
- Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cabo Verde, Canada, Chile, China, Costa Rica, Croatia, Czechia, Denmark,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1025-2026, Z-1026-2026
- Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit
- Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Ongoing Class II Medical devices
- Recall started
- 10 November 2025
- Reason
- Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Can
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1221-2026
- COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter
Ongoing Class II Medical devices
- Recall started
- 7 November 2025
- Reason
- It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
- Distribution
- US Nationwide. Global Distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0889-2026
- Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901
Ongoing Class II Medical devices
- Recall started
- 1 October 2025
- Reason
- An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
- Distribution
- Worldwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0428-2026, Z-0429-2026
- Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
- Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
Ongoing Class II Medical devices
- Recall started
- 22 September 2025
- Reason
- Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, CT, FL, IL, NY, OH, OR, TN, and TX. The country of India.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0541-2026
- DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent detection channels us
Ongoing Class II Medical devices
- Recall started
- 19 September 2025
- Reason
- Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0216-2026, Z-0215-2026
- DxI 9000 Access Immunoassay Analyzer, Part Number C11137
- DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Module, DxC 500i (Part Number C13252)
Ongoing Class II Medical devices
- Recall started
- 17 September 2025
- Reason
- Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, CA, FL, GA, HI, IA, ID, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, PR, QC, SC, TX, WA, WV and the country of Canada.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0441-2026
- SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706
Ongoing Class II Medical devices
- Recall started
- 8 September 2025
- Reason
- Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0412-2026
- CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Ongoing Class II Medical devices
- Recall started
- 8 September 2025
- Reason
- Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
- Distribution
- International distribution to the country of Poland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1794-2026
- MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan pa
Ongoing Class II Medical devices
- Recall started
- 3 September 2025
- Reason
- Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
- Distribution
- US: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico Singapore
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0071-2026
- SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
Ongoing Class II Medical devices
- Recall started
- 13 August 2025
- Reason
- Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Br
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2574-2025, Z-2572-2025, Z-2573-2025, Z-2575-2025
- UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322
- UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478
- UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
- UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
Ongoing Class II Medical devices
- Recall started
- 4 August 2025
- Reason
- On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.
- Distribution
- Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0032-2026
- DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer
Ongoing Class II Medical devices
- Recall started
- 24 July 2025
- Reason
- Due to the likely presence of contamination in well(s).
- Distribution
- U.S. Nationwide distribution in the states of CA, CO, IN, MA, MD, MN, MO, NC, NJ, OK, PA, and TX.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1609-2026
- MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultativ
Ongoing Class II Medical devices
- Recall started
- 10 July 2025
- Reason
- Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer.
- Distribution
- Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy, Malaysia, Mexico, Philippines, Spain.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2322-2025
- Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860
Ongoing Class II Medical devices
- Recall started
- 26 June 2025
- Reason
- Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
- Distribution
- US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hun
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2264-2025
- UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,
Ongoing Class II Medical devices
- Recall started
- 25 June 2025
- Reason
- Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
- Distribution
- Worldwide distribution - US Nationwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2267-2025, Z-2271-2025, Z-2268-2025, Z-2270-2025, Z-2269-2025, Z-2266-2025
- DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
- DxI 600 Access Immunoassay Analyzer W/Spot B, Part Number A71460
- DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456
- DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461
- UniCel DxI 600 Access Immunoassay Analyzer, Part Number A30260
- UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100
Ongoing Class II Medical devices
- Recall started
- 25 June 2025
- Reason
- Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
- Distribution
- International distribution in the country of Japan.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1772-2026
- MicroScan Neg MIC 3J REF C54814
Ongoing Class II Medical devices
- Recall started
- 12 June 2025
- Reason
- A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
- Distribution
- US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2152-2025
- Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.
Ongoing Class II Medical devices
- Recall started
- 7 May 2025
- Reason
- Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI, ID, IL, IN, KY, LA, ME, MI, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, WA, WI and the countries of Algeria, Australia, Austria, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, France, Germany, Hungary, India, Ireland, Israel,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1996-2025
- DxI 9000 Access Immunoassay Analyzer C11137
Ongoing Class II Medical devices
- Recall started
- 5 May 2025
- Reason
- Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
- Distribution
- US State: FL. UAE, Vietnam
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1901-2025
- Access Testosterone assay, Catalog Number 33560
Ongoing Class II Medical devices
- Recall started
- 23 April 2025
- Reason
- A new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Kazakhstan, Spain, Germany, Italy, Belgium, Croatia, Taiwan, Province of China, Poland, India, Singapore, Switzerland, France, Libya, United Kingdom of Great Britain and Northern Irland, Oman, Algeria, Czechia, Korea, Public o
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1913-2025
- DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software version 1.20
Ongoing Class II Medical devices
- Recall started
- 23 April 2025
- Reason
- A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
- Distribution
- Worldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries of: Italy, Germany, Spain, New Zealand, France, Australia, Slovakia, Switzerland, Ireland, Brazil, Taiwan, Province of China, India, United Kingdom of Great Britain and Northern Ireland, Hungary, Austria, Portugal, Croatia
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1926-2025
- DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software versions 1.20
Ongoing Class II Medical devices
- Recall started
- 8 April 2025
- Reason
- Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Australi
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2028-2025
- REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Dat
Ongoing Class II Medical devices
- Recall started
- 27 March 2025
- Reason
- When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1664-2025
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer
Ongoing Class II Medical devices
- Recall started
- 20 March 2025
- Reason
- Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.
- Distribution
- U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI O.U.S.: Australia, Belgium, Bosnia and Herzegovina, Canada, Chile, Croatia, Egypt, Ethiopia, France, Germany, Ghana, India, Italy, Korea, Lebanon, Libya, Malaysia, Netherlands, North Macedonia, Oman, Paraguay, Philippines,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1784-2025
- DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that
Ongoing Class II Medical devices
- Recall started
- 17 March 2025
- Reason
- The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, CA, CO, GA, ID, IL, IN, LA, MA, MI,MN, MS, MT, NC, NM, NV, NY, OK, OR, PA, PR, SC, TN, TX, WA, and WI. The countries of Australia, Belgium, Bosnia and Herzegovina, Canada, Chile, Croatia, Egypt, Ethiopia, France, Georgia, German
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1672-2025
- DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 Software versions: V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3 The Beckman Coulter DxC 500 AU Clinical Che
Completed Class II Medical devices
- Recall started
- 27 February 2025
- Reason
- Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a delayed or improper diagnosis and/or repeat testing.
- Distribution
- International distribution to the country of Andorra, Bulgaria, Denmark, Estonia, France, Germany, Greece, Ireland, Italy, Maldives, Netherlands, Portugal, Romania, Spain, Switzerland, Turkey, United Kingdom of Great Britain and Northern Ireland
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2115-2025
- Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850
Ongoing Class II Medical devices
- Recall started
- 26 February 2025
- Reason
- Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the operator at the analyzer or automatically triggered by predefined rules in System Configuration, the calculated result will not be recalculated using the constituent tests rerun results. The system will report only the calculated result using the initial constituent tests results to the operator and LIS/Remisol. The issue was identified by Beckman Coulter internal engineering testing. The defect may cause a delay in reporting patient results.
- Distribution
- Worldwide Distribution: U.S. (nationwide) to states of: AL, CA, CO, GA, ID, IL, IN, LA, MA, AI, MN, MS, MT, NC, NM, NV, NY, OK, OR, PA, PR, SC, TN, TX, WA, and WI; O.U.S. (foreign) to countries of: Australia, Belgium, Bosnia/Herzegovina, Canada, Croatia, Egypt, Ethiopia, France, Germany, Ghana, Indi
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1517-2025
- DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that m
Ongoing Class II Medical devices
- Recall started
- 6 February 2025
- Reason
- Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, TX, WA.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1387-2025
- DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278
Ongoing Class II Medical devices
- Recall started
- 5 February 2025
- Reason
- Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1386-2025
- Access 2 Immunoassay Analyzer, Catalog Number 81600N
Ongoing Class II Medical devices
- Recall started
- 5 February 2025
- Reason
- the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
- Distribution
- US Nationwide. Global Distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1388-2025
- DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.
Ongoing Class II Medical devices
- Recall started
- 3 February 2025
- Reason
- Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.
- Distribution
- Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, P
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1249-2025
- Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal pr
Ongoing Class II Medical devices
- Recall started
- 2 January 2025
- Reason
- Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philipp
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1112-2025
- Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems
Ongoing Class II Medical devices
- Recall started
- 26 November 2024
- Reason
- Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0803-2025
- Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoas
Ongoing Class II Medical devices
- Recall started
- 18 November 2024
- Reason
- Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.
- Distribution
- Worldwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0727-2025
- Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.
Ongoing Class II Medical devices
- Recall started
- 13 November 2024
- Reason
- The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results
- Distribution
- US Nationwide. Global Distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1326-2025
- DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137
Ongoing Class II Medical devices
- Recall started
- 8 November 2024
- Reason
- The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
- Distribution
- Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, Egypt, France, Germany, Ghana, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, S. Korea, Spain, Sw
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0673-2025
- Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
Ongoing Class II Medical devices
- Recall started
- 30 September 2024
- Reason
- The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0313-2025
- The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Sy
Ongoing Class II Medical devices
- Recall started
- 5 September 2024
- Reason
- IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.
- Distribution
- Distribution was made nationwide to the U.S., including Puerto Rico. There was government distribution but no military distribution. Foreign distribution was made to Mexico.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0265-2025
- Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody II (TgAb) assay is used for the quantitative determination of thyroglo
Ongoing Class II Medical devices
- Recall started
- 28 August 2024
- Reason
- Certain lots of Access TSH reagent packs may produce erroneously low TSH (thyroid-stimulating hormone) results.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Croatia, France, Germany, Ireland, Italy, Slovakia, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0094-2025
- The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative etermdination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma usi
Ongoing Class II Medical devices
- Recall started
- 31 July 2024
- Reason
- Firm identified two lots of reagents that are not meeting the sensitivity statement in the IFU
- Distribution
- U.S.: AZ, CA, CO, FL, GA, HI, IL, MA, MI, MO, ND, NY, OH, PA, and TX O.U.S.: Australia, Belgium, Brazil, Chile, China, Czechia, Ecuador, France, Germany, Greece, Hong Kong, India, Italy, Kuwait, Mexico, Philippines, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan Province of China
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0118-2025
- IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent REF 447610 IGM reagent, when used in conjunction with IMMAGE Immunochemistry Systems and Calibrator 1, is intended for the quantitative determination of immu
Ongoing Class III Medical devices
- Recall started
- 16 July 2024
- Reason
- Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.
- Distribution
- worldwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0093-2025
- Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.
Ongoing Class II Medical devices
- Recall started
- 12 July 2024
- Reason
- Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.
- Distribution
- OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Africa, Spain, and United Kingdom of Great Britain and Northern Ireland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1529-2025
- DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body
Ongoing Class II Medical devices
- Recall started
- 10 July 2024
- Reason
- Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United K
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-3180-2024
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument
Ongoing Class II Medical devices
- Recall started
- 14 June 2024
- Reason
- Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
- Distribution
- Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2545-2024
- Discrete photometric chemistry analyzer for clinical use.
Ongoing Class II Medical devices
- Recall started
- 4 June 2024
- Reason
- The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2414-2024
- Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Ongoing Class II Medical devices
- Recall started
- 27 May 2024
- Reason
- LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Albania, Algeria, Argentina, Australia,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2428-2024
- Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
Ongoing Class II Medical devices
- Recall started
- 22 May 2024
- Reason
- When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2735-2024
- DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Ongoing Class II Medical devices
- Recall started
- 22 May 2024
- Reason
- The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, and TX. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Port
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2544-2024
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Ongoing Class II Medical devices
- Recall started
- 18 April 2024
- Reason
- Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids, by correction. The reason for the recall is: Beckman Coulter has determined that the DxI 9000 Access Immunoassay analyzer accepts calibrator values scanned from multiple calibrator cards when an operator configures a new calibrator lot. The analyzer accepts the calibrator lot values from any scanned calibrator card, regardless of the assay test code or calibrator material lot number. The software is not validating the integrity of each calibrator barcode scanned to ensure they are all from the same calibrator card. This includes all software versions and all DxI 9000 instruments that are currently in the field. This issue does not impact calibrator material with 2D calibrator card barcodes. This may cause a failed calibration curve. The health risk to patients is dependent upon the analytes that could be impacted by the error, to include receiving delayed or erroneous results. The issue was identified by Beckman Coulter while conducting a field action on DxI 9000 Access Immunoassay Analyzers.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2264-2024
- DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic
Ongoing Class II Medical devices
- Recall started
- 18 April 2024
- Reason
- When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.
- Distribution
- Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0025-2025
- Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human
Ongoing Class II Medical devices
- Recall started
- 5 March 2024
- Reason
- Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
- Distribution
- Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1546-2024
- DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Ongoing Class II Medical devices
- Recall started
- 29 February 2024
- Reason
- The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000 Access Immunoassay Analyzer may generate a System Event log message that indicates PnP errors. Any cancelled tests will have an associated SYS flag. The analyzer may enter the Red-system status because of this issue. The analyzer also may not eject RVs, where unbound conjugate could potentially splash up on the sidewall of the RV, not being properly washed away, resulting in an artificially higher signal (RLU). The analyzer may not detect this splashing and continue with normal operation, which can cause false low or high erroneous results. Consequently, the analyzer may cancel in-progress tests. These errors could potentially cause delay reporting patient test results and patient treatments. The degree of the potential health hazard is dependent on the analytes that are potentially delayed, especially those analytes that would be used in the acute setting when treatment decisions are immediate. In this particular recall the highest risk analyte is troponin. Health consequences could include permanent impairment or possibly death if medical treatment decisions are delayed due to a delay in obtaining patient results. The issue was identified by Beckman Coulter Inc. manufacturing staff at their location in Brea California on February 29, 2024, during in-process acceptance inspection activities.
- Distribution
- Worldwide - US Nationwide distribution in the state of Indiana and the countries of Italy, Sweden, Ireland, Spain.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2237-2024
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. It is an in-vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determ
Ongoing Class II Medical devices
- Recall started
- 8 February 2024
- Reason
- Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
- Distribution
- Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, P
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1303-2024
- Access Total T4, REF 33800, IVD, thyroxine reagent assay.
Ongoing Class II Medical devices
- Recall started
- 7 February 2024
- Reason
- On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman Coulter has identified that Access Substrate bottles in the identified lots were released from manufacturing without labels. Missing label may cause some patient results reporting delay but will not cause any errorreous results. However, because labs are required to have redundancy such that delay would not significantly impact clinical management and because the defect is easily recognized by the user, the probability of harm is remote.
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Denmark, Ecuador, Egypt, El Salvador, Es
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1376-2024
- Access Substrate, REF 81906, For use with the Access Immunoassay Systems
Completed Class II Medical devices
- Recall started
- 3 January 2024
- Reason
- A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1115-2025
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Ongoing Class II Medical devices
- Recall started
- 19 December 2023
- Reason
- Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
- Distribution
- US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0784-2024, Z-0785-2024, Z-0787-2024, Z-0786-2024
- Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
- iQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e) C10683, f) 700-3375, g) 700-3325, h) C10684
- iQ Waste Well Adapter, Part Number: 700-3393
- REV 2 Flowcell Capsule Spare, Part Number: C04154
Ongoing Class II Medical devices
- Recall started
- 21 November 2023
- Reason
- There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Croatia, Czech Republic, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0981-2024, Z-0982-2024
- Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410
- DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,
Ongoing Class II Medical devices
- Recall started
- 12 September 2023
- Reason
- The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.
- Distribution
- Worldwide - US Nationwide distribution in the state of Florida, and the countries of Croatia, Czechia, France, Germany, Ireland, Italy, Netherlands, New Zealand, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0107-2024
- DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic
Ongoing Class II Medical devices
- Recall started
- 12 July 2023
- Reason
- Lipemic interference failed to meet the performance specification listed within the IFU.
- Distribution
- US: AL AR AZ CA CO CT DC DE FL GA IA IL IN KS KY LA MA MD MI MN MO MS NC ND NE NJ NM NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV OUS: Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, C¿te d'Ivoire, Croatia, Cur
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2488-2023
- Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2
Ongoing Class II Medical devices
- Recall started
- 6 July 2023
- Reason
- It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.
- Distribution
- worldwide except US, Canada, and China
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0100-2024
- Access Thyroglobulin Calibrators
Ongoing Class II Medical devices
- Recall started
- 20 June 2023
- Reason
- Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2237-2023
- Ringed DxI Reaction Vessels (RVs)
Ongoing Class II Medical devices
- Recall started
- 31 May 2023
- Reason
- Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2138-2023
- COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells
Ongoing Class II Medical devices
- Recall started
- 17 May 2023
- Reason
- Reagent lot numbers include an insufficient concentration of blocking reagent.
- Distribution
- Distribution was made to CA, CO, CT, FL, GA, IL, IN, MA, MI, NC, NE, NJ, NY, OH, SC, TN, TX, UT, and WA. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, China, Czech Republic, France, Germany, Hong Kong, India, I
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2043-2023
- Access Hybritech p2PSA reagent, REF B03704 and REF A49752.
Ongoing Class II Medical devices
- Recall started
- 5 April 2023
- Reason
- Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.
- Distribution
- Worldwide Distribution
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1598-2023, Z-1596-2023, Z-1599-2023, Z-1597-2023
- DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
- iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Numb
- UA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis
- DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis
Ongoing Class II Medical devices
- Recall started
- 5 April 2023
- Reason
- There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
- Distribution
- Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Guam, & Puerto Rico and OUS (foreign) countries of: Brazil
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1606-2023, Z-1608-2023, Z-1607-2023
- Beckman Coulter Bicarbonate, REF: OSR6137, 4x25 mL
- Beckman Coulter Bicarbonate, REF: OS6637, 4x173 mL
- Beckman Coulter Bicarbonate, REF: OSR6237, 4x50 mL
Ongoing Class II Medical devices
- Recall started
- 15 March 2023
- Reason
- The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not being detected by QC with the potential that clinical correlation with other thyroid function tests will be inconsistent requiring additional testing; however the clinical assessment and treatment decisions are not made in isolation on a single test result especially, in the setting of thyroid function testing. False positive test results could lead to a euthyroid patient being interpreted clinically as having elevated thyroid function (hyperthyroid) or a hypothyroid patient being within the normal range. Such misdiagnoses may lead to inappropriate treatment of a euthyroid patient with anti-thyroid medications or hypothyroid patient not getting sufficient treatment since their free T3 test result demonstrates them to be in the euthyroid range.
- Distribution
- Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Brazil, Canada, China, Costa Rica, El Salvador, Georgia, Hong Kong, Malaysia, Mexico, Myanmar, New Zealand, Peru, Philippines, Qatar, Singapore, Taiwan, Thailand, Uruguay, and Viet Nam.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1419-2023
- Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.
Ongoing Class II Medical devices
- Recall started
- 26 January 2023
- Reason
- An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period.
- Distribution
- Worldwide and Nationwide Distribution. There was government/military distribution. Foreign distribution was made to Algeria, Andorra, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Bonaire, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croati
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1257-2023
- Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, Dx
Ongoing Class II Medical devices
- Recall started
- 18 November 2022
- Reason
- Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN 800-7212
- Distribution
- US and Canada, Brazil, Korea, Mexico, Puerto Rico, Taiwan
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0777-2023
- BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.
Ongoing Class II Medical devices
- Recall started
- 3 October 2022
- Reason
- Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of AK, FL, IA, ID, MI MS, MT, NC, NE, PA, SC, TX WA, WI, and WV. The countries of Chile, Germany, Indonesia, Italy, Japan, Mexico, Peru, Portugal, South Africa, Spain, United Kingdom of Great Britain and Northern Ireland, and Uruguay.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0269-2023
- MicroScan autoSCAN-4 Instrument, Catalog Number B1018-280
Ongoing Class III Medical devices
- Recall started
- 16 February 2022
- Reason
- IRISpec CA failures for glucose.
- Distribution
- Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Brazil, Canada, Curacao, Guatemala, Mexico, Philippines, Singapore, and Taiwan.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0984-2022
- Beckman Coulter IRISpec CA/CB/CC control, Catalog #800-7211, containing 3 x 100 mL bottles each of CA Control, CB Control, and CC Control.
Terminated Class II Medical devices
- Recall started
- 8 December 2021
- Terminated
- 28 January 2025
- Reason
- Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
- Distribution
- US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Mont
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0692-2022
- AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850
Ongoing Class II Medical devices
- Recall started
- 10 November 2021
- Reason
- Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.
- Distribution
- US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0399-2022
- SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825
Ongoing Class II Medical devices
- Recall started
- 9 August 2021
- Reason
- Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.
- Distribution
- U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herz
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0032-2022
- Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK
Terminated Class II Medical devices
- Recall started
- 5 August 2021
- Terminated
- 1 June 2023
- Reason
- Firm's submission for Emergency Use Authorization (EUA) contained insufficient evidence to be granted EUA Authorization to distribute
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, GA, IL, IA, KY, MI, NJ, NY, NC, OH, PA, Puerto Rico, TN, TX, VA, and WA.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2508-2021
- Access SARS CoV-2 IgG (1st IS) Assay: Access Immunoassay Systems REAGENT PACK, REF C74339 Access SARS-CoV-2 IgG (1st IS) Calibrators, REF C74340 Access SARS-CoV-2 (1st IS) Quality Control (QC), REF C74341
Ongoing Class II Medical devices
- Recall started
- 28 July 2021
- Reason
- Missing splash guard which may expose users to hazardous materials. .
- Distribution
- U.S. Nationwide Distribution in the states of: CA, GA, IA, ID, IL, IN, LA, MA, MI, MS, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WA, and WV O.U.S.: Canada, China, Spain, France, United Kingdom, Israel, Italy, South Korea, Kuwait, Lebanon, Mexico, Malaysia, Netherlands, Portugal, Qatar, Saudi Arabia, Thai
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0117-2022
- Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116
Terminated Class II Medical devices
- Recall started
- 21 July 2021
- Terminated
- 10 October 2024
- Reason
- There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).
- Distribution
- US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH, PA, TX, WV.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2602-2021
- DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118
Terminated Class II Medical devices
- Recall started
- 15 July 2021
- Terminated
- 4 June 2024
- Reason
- There is a potential that the data management system may add additional cells to the patient request which could lead to erroneous patient result.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Al, CA, FL, GA, IL, IN, MA, NC, OH, PA, TN, TX and the countries of Australia, Belgium, Canada, China, Czech Republic, France, Italy, Netherlands, New Zealand, Poland, Russia, Switzerland, Taiwan, United Kingdom, United States, Zim
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2332-2021
- Normand Remisol Advance Data Manager, REF: C69412, C69413, C44703, and C57017 is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments.
Terminated Class II Medical devices
- Recall started
- 1 July 2021
- Terminated
- 14 June 2023
- Reason
- SARS-CoV-2 Antigen Assay test may generate false positive results which could result in incorrect diagnosis .
- Distribution
- US Nationwide Distribution in the states of Alabama Alaska Arizona Arkansas California Connecticut Delaware Florida Georgia Idaho Illinois Indiana Iowa Kentucky Louisiana Maine Maryland Massachusetts Michigan Mississippi New Hampshire New Jersey New York North Carolina Ohio Oklahoma Oregon Pennsylv
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2257-2021, Z-2258-2021, Z-2255-2021, Z-2256-2021
- ACCESS SARS-CoV-2 ANTIGEN QC, REF: C68670, CONTENTS: 3 X 6.0 mL QC1, 3 x 6.0 mL QC2, 1 QC VALUE CARD, IVD, Rx Only, UDI: (01)15099590742737;
- ACCESS SARS-CoV-2 ANTIGEN EXTRACTION SOLUTION, CONTENTS: 6 x 6.0 mL EXTRACTION SOLN, IVD, Rx Only, UDI: (01)15099590743017
- ACCESS SARS-CoV-2 Antigen , REF: C68668, IVD, Rx only, UDI: (01) 15099590742713;
- Access SARS-CoV-2 Antigen Calibrators, Catalog, REF: C68669, Contents: 1x2.5 mL C0, 1x2.5mL C1, 1x CAL CARD, UDI: (01)15099590742720;
Terminated Class II Medical devices
- Recall started
- 23 June 2021
- Terminated
- 2 July 2026
- Reason
- Software anomalies that may lead to the generation of erroneous results.
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, and Hawaii. The countries of Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, C
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2241-2021
- Kaluza C Flow Cytometry Software Versions: 1.0*, 1.1.1 and 1.1.2 Kits: C10574 - Kaluza C Single Perpetual License C10575 - Kaluza C Single User 1-year License C10576 - Kaluza C 10 User Network License C10577 - Kalu
Terminated Class II Medical devices
- Recall started
- 7 May 2021
- Terminated
- 27 March 2023
- Reason
- SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.
- Distribution
- US: Alabama, Alaska, Arizona, Arkansas, California, Florida, Georgia, Illinois, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Tennessee
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1905-2021, Z-1906-2021
- Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744
- Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751
Terminated Class II Medical devices
- Recall started
- 14 April 2021
- Terminated
- 19 May 2023
- Reason
- specimens collected in transport media may result is false negative results when stored in certain conditions.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Alabama, Arizona, Arkansas, California, Connecticut, Delaware, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, New Hampshire, New Jersey, New York, North
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1670-2021
- Access SARS-CoV-2 Antigen, REF C68668, UDI: (01)15099590742713, 3.30 mL/R1a, 4.70 mL/R1b, 4.20 mL/R1c, 3.10 mL, R1d, 3.10 mL, R1e
Terminated Class II Medical devices
- Recall started
- 30 December 2020
- Terminated
- 19 September 2022
- Reason
- The Quality Control (QC) card has a label error for the standard deviation (SD) values for one out of the two levels of QC.
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, GA, IA, ID, IL, IN, KS,KY, LA, MD, MI, MN, MO, NC, ND, NY, OH, OK, OR, PA, Puerto Rico, SD, TX, WA, and WV. The countries of Andorra, Belgium, Brazil, Bulgaria, Canada*, Croatia, France, Germany, Hong Ko
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0875-2021
- ACCESS SARS-CoV-2 IgG QC SARS-CoV-2 IgG REF C58964 - Product Usage: use on the Access Family of Immunoassay Systems only.
Terminated Class II Medical devices
- Recall started
- 2 December 2020
- Terminated
- 3 January 2023
- Reason
- Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.
- Distribution
- U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Re
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0720-2021
- MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some
Terminated Class II Medical devices
- Recall started
- 26 August 2020
- Terminated
- 24 May 2024
- Reason
- Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, WA, WI and Puerto Rico. The countries of Australia, Belgium, Bosnia and Herzegovina, Brazil, Can
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-3006-2020
- BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.
Terminated Class II Medical devices
- Recall started
- 18 August 2020
- Terminated
- 8 September 2021
- Reason
- The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of CA, FL and NJ. The countries of Algeria, Egypt, Philippines and Taiwan.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-3028-2020
- UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage: System reagent for the quantitative determination of Unsaturated Iron Binding Capacity (UIBC) in hu
Terminated Class II Medical devices
- Recall started
- 30 March 2020
- Terminated
- 31 August 2023
- Reason
- A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
- Distribution
- US Distribution: Nationwide Foreign Distribution: Worldwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1944-2020, Z-1945-2020
- iChem¿VELOCITY Urine Chemistry System-
- iQ200 Series Urine Microscopy Analyzer- All part numbers
Terminated Class II Medical devices
- Recall started
- 17 February 2020
- Terminated
- 13 May 2021
- Reason
- The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
- Distribution
- Distribution in US including Puerto Rico, Canada, and Taiwan
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1535-2020
- iChemVELOCITY Strips, Catalog 800-7212
Terminated Class II Medical devices
- Recall started
- 29 January 2020
- Terminated
- 28 April 2021
- Reason
- Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries of CANADA, JAPAN, LEBANON, SAUDI ARABIA, TAIWAN.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2136-2020
- ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.
Terminated Class II Medical devices
- Recall started
- 17 January 2020
- Terminated
- 27 April 2021
- Reason
- A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases some customers reported that Data Management System computers would not restart. If the computer is unable to restart, then the system cannot download data generated from MicroScan instruments or manually entered microbiology test results, resulting in a potential of delayed results.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Wisconsin, New York, Delaware, Indiana, Ohio, Tennessee, Florida, Georgia, Illinois, South Carolina, California, New Jersey, Alabama, Massachusetts, Michigan, Texas, West Virginia, Wyoming, Missouri, New Mexico, North Carolina, Min
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1673-2020
- LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimic
Terminated Class II Medical devices
- Recall started
- 13 December 2019
- Terminated
- 30 April 2021
- Reason
- Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MI, MS, NE, NY, NC, ND, OH, OK, PA, SD, TN, TX, UT, WV, WI, WY and countries of Canada, China, Japan, Paraguay.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1674-2020
- Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial suscept
Terminated Class II Medical devices
- Recall started
- 7 November 2019
- Terminated
- 11 September 2023
- Reason
- Beckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag S3 and may result in erroneous Specific Gravity (SG) results.
- Distribution
- Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, France,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0843-2020
- iChem VELOCITY Urine Chemistry Analyzer
Terminated Class II Medical devices
- Recall started
- 28 October 2019
- Terminated
- 1 April 2026
- Reason
- Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).
- Distribution
- Worldwide distributions.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0913-2020, Z-0914-2020
- iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700-7176-001, 700-7177-001, B75502 - Product Usage: used to automate the urine chemistry analysis profile using iChem¿ VELOCITY Urine Chemistry Strips.
- iQ200 Series Urine Microscopy Analyzer, Catalog numbers: 700-3345, 700-3375, 700-3325, 700-3347, 700-3300RFB, 700- 3320, 700-3320RFB, 700-3322, 700-3322RFB, 700-3326, 700- 3340, 700-3370, B48999, C10683, C10684 - Produc
Ongoing Class II Medical devices
- Recall started
- 21 October 2019
- Reason
- WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1083-2022, Z-1086-2022, Z-1085-2022, Z-1087-2022, Z-1088-2022, Z-1084-2022
- MicroScan WalkAway-40 plus Instrument REF B1018-283
- DxM 1096 MicroScan WalkAway Instrument REF B1018-496
- DxM 1040 MicroScan WalkAway Instrument REF B1018-440
- MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R
- MicroScan WalkAway-96 plus Reconditioned Instrument REF B1018-396R
- MicroScan WalkAway-96 plus Instrument REF B1018-284
Terminated Class II Medical devices
- Recall started
- 7 October 2019
- Terminated
- 27 April 2021
- Reason
- When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0516-2020
- UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining de
Terminated Class II Medical devices
- Recall started
- 4 September 2019
- Terminated
- 28 April 2021
- Reason
- Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
- Distribution
- Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1751-2020, Z-1752-2020
- Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators. The assay provide
- Opiate 300 Ng Reagent (OP 1 X 250) for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems OP 300 Urine Calibrators, Catalogue No.475024. The OP assay provides a rapid screening procedure for determ
Terminated Class II Medical devices
- Recall started
- 4 September 2019
- Terminated
- 11 January 2023
- Reason
- Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Distribution
- U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1370-2021, Z-1372-2021, Z-1373-2021, Z-1371-2021
- MicroScan WalkAway-40 plus Instrument
- MicroScan WalkAway-40 plus Reconditioned Instrument
- MicroScan WalkAway-96 plus Reconditioned Instrument
- MicroScan WalkAway-96 plus Instrument
Terminated Class II Medical devices
- Recall started
- 15 August 2019
- Terminated
- 28 February 2024
- Reason
- It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.
- Distribution
- Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Je
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1423-2020, Z-1422-2020, Z-1424-2020, Z-1426-2020, Z-1425-2020
- Triglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233228
- Enzymatic Creatinine Reagent (CR-E 2 X 200), REF A60298, For in Vitro Diagnostic Use, Rx Only, UDI: 15099590625733
- Uric Acid Reagent (URIC 2 X 300), REF: 442785, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590575281
- Total Bilirubin Reagent (TBIL 2 x 300 ) and (TBIL 2 x 400), Ref:442745 (300 tests/cartridge) and 476861 (400 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233075 (300 tests/cartridge) and 150995906
- Direct Bilirubin Reagents (DBIL 2 x 200 ) and (DBIL 2 x 300), Ref:439715 (200 tests/cartridge) and 476856 (300 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590575120(200 tests/cartridge) and 1509959
Terminated Class II Medical devices
- Recall started
- 12 June 2019
- Terminated
- 11 December 2020
- Reason
- The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may result in false positive or false negative results.
- Distribution
- Worldwide distribution - US Nationwide distribution and to multiple countries.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1799-2020
- SYNCHRON Systems Barbiturates Reagent (BARB 1 X 250); Catalog No. 475012 - Product Usage: BARB reagent, in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators
Terminated Class II Medical devices
- Recall started
- 17 May 2019
- Terminated
- 11 March 2022
- Reason
- Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.
- Distribution
- Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1741-2020
- Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.
Terminated Class II Medical devices
- Recall started
- 30 April 2019
- Terminated
- 28 December 2021
- Reason
- The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
- Distribution
- US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusett, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2239-2020, Z-2241-2020, Z-2240-2020, Z-2238-2020, Z-2242-2020, Z-2237-2020
- Beckman Coulter ACCESS Immunoassay Systems, Access Total T3 Triiodothyronine , REF 33830 (UDI: 15099590227210) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800,
- Beckman Coulter ACCESS Immunoassay Systems, Access GI Monitor Cancer Antigen 19-9, REF 387687 ( UDI: 15099590231842) the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel
- Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600
- Beckman Coulter Access Immunoassay Systems, Access Free T3 Triiodothyronine, REF A13422 (UDI: 15099590201661) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, U
- Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800,
- Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i
Ongoing Class II Medical devices
- Recall started
- 5 December 2018
- Reason
- When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
- Distribution
- US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Morocco, Netherlands, Nigeria, Poland, Portugal, Qatar, Saudi Ara
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1078-2019, Z-1077-2019
- UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. Sy
- The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A714
Ongoing Class II Medical devices
- Recall started
- 9 November 2018
- Reason
- Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .
- Distribution
- US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0492-2020
- AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255
Terminated Class II Medical devices
- Recall started
- 18 October 2018
- Terminated
- 10 November 2020
- Reason
- Falsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction.
- Distribution
- Distribution was nationwide, including Puerto Rico. There was also military/government distribution. Foreign distribution was made to Canada, Argentina, Brazil, Chile, Colombia, Mexico, Peru, Philippines, Singapore, and Taiwan.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1937-2019
- iChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland, Made in Germany.
Terminated Class II Medical devices
- Recall started
- 18 September 2018
- Terminated
- 21 February 2024
- Reason
- Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2763-2019
- The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanit
Terminated Class II Medical devices
- Recall started
- 17 September 2018
- Reason
- Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.
- Distribution
- US: CA and NY OUS: United Kingdom, Switzerland and Germany
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1982-2019
- Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A) Kaluza C Perpetual License (B) Kaluza C Single Use 1 Year License (C) Kaluza C 5 User Network
Terminated Class II Medical devices
- Recall started
- 24 August 2018
- Terminated
- 2 August 2024
- Reason
- NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.
- Distribution
- Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0210-2020, Z-0211-2020
- Beckman Coulter Lipase, REF OSR6130 and OSR6230.
- Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698.
Terminated Class I Medical devices
- Recall started
- 30 July 2018
- Terminated
- 31 March 2022
- Reason
- Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
- Distribution
- Worldwide Distribution - US Nationwide including Puerto Rico. There was government and military distribution. Foreign distribution was made to Andorra, Australia, Austria, Belgium, Canada, China, Croatia, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Korea, Kuwait, Latv
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1382-2019, Z-1383-2019, Z-1384-2019
- UniCel DxH 800 Coulter Cellular Analysis System, REF numbers 629029, B24465, B24802, B68304, B66445, and B63322. Product Usage: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology anal
- UniCel DxH 600 Coulter Cellular Analysis System, REF number B23858. Product Usage: Note: The DxH 600 hematology analyzer is the same instrument as the DxH 800 (K140911), except that it is a benchtop unit with no inst
- UniCel DxH 900 Coulter Cellular Analysis System, REF number C11478 Product Usage: Note: The DxH 900 hematology analyzer is CLIA categorized under document CR180294 and is currently marketed as a member of the instrum
Terminated Class III Medical devices
- Recall started
- 24 May 2018
- Terminated
- 8 December 2022
- Reason
- Sorting-Drive software version 4.2 (released in October 2015) introduced a new option which included the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of Korea, Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0354-2019, Z-0358-2019, Z-0353-2019, Z-0359-2019, Z-0356-2019, Z-0355-2019, Z-0357-2019
- Software Package, Sorting-Drive 4.2 User Kit, Part No. B73488, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 Family of products.
- Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman Coulter, Inc. The Sorting-Drive Plus Kit includes the Sorting-Drive 4.2.1 User Kit which includes the Sorti
- Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking
- Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes the Sorting-Drive 4.2
- Software Package, iPAW 4.2 User Kit, Part No. B73489, Manufactured by: Beckman Coulter Biomedical GmbH The iPAW 4.2 User Kit (PN B73489) includes the iPAW software application CD (version 4.2). This function is prima
- Software Package, Sorting-Drive 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95347, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Aut
- Software Package, iPAW 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95343, Manufactured by: Beckman Coulter Biomedical GmbH Sorting-Drive software version 4.2 (released in October 2015) introduced a new opt
Terminated Class II Medical devices
- Recall started
- 4 May 2018
- Terminated
- 7 August 2020
- Reason
- The drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.
- Distribution
- NJ, TX, IA, PA, FL, NE, OH, NY, CA, MO, OK, LA
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2599-2018
- MicroScan Pos Combo Panel Type 43, Catalog number B1017-217for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive coc
Terminated Class II Medical devices
- Recall started
- 29 March 2018
- Terminated
- 27 June 2019
- Reason
- The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.
- Distribution
- US Distribution to states of: AL, IA, GA, and FL.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1894-2018
- iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an
Terminated Class II Medical devices
- Recall started
- 26 March 2018
- Terminated
- 19 December 2018
- Reason
- Beckman Coulter has received customer complaints of lowered minimum inhibitory concentration (MIC) for antimicrobial agents Cefaclor (Ctr) and Penicillin (P) with control organism Streptococcus pneumoniae (ATCC 49619) during quality control (QC) testing on MICroSTREP plus(R) Panel Type 1 hydrated with Mueller-Hinton Broth with 3% lysed horse blood part number B1015-25 lot number 1005986140.
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Canada, Kuwait, South Korea and Saudi Arabia.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1804-2018
- BECKMAN COULTER MicroScan MICroSTREP Plus Panel, MicroScan Mueller-Hinton Broth with 3% lysed horse blood, Catalog Number 81015 MicroScan Mueller Hinton Broth with 3% Lysed Horse Blood is used for inoculation of Micro
Terminated Class II Medical devices
- Recall started
- 2 March 2018
- Terminated
- 27 July 2023
- Reason
- Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1643-2018
- BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167, 800-7719, 800-7720, 800- 7106, 800-3049, 800-3050, 800-3079, 800-3080, 800-7190, 800-7713, 800-7714, 800-7715, 700- 7176-001
Terminated Class II Medical devices
- Recall started
- 2 March 2018
- Terminated
- 6 July 2023
- Reason
- Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.
- Distribution
- Worldwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1362-2018
- BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370
Terminated Class II Medical devices
- Recall started
- 29 January 2018
- Terminated
- 27 June 2019
- Reason
- The device may process the same sample with two different sample IDs and sample information while using the single tube loader, which has the potential for erroneous results due to the mis-identification.
- Distribution
- Distribution was made nationwide in the U.S. and in Puerto Rico. There was also government distribution, but no military distribution. Foreign distribution was made to Angola, Australia, Belgium, Botswana, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ghan
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1341-2018
- AQUIOS CL Flow Cytometer System, Catalog #B30166. The AQUIOS CL Flow Cytometer system is an automated analyzer that use a no-wash sample preparation process.The firm name on the label is Beckman Coulter Ireland, Inc., C
Completed Class II Medical devices
- Recall started
- 19 January 2018
- Reason
- Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.
- Distribution
- Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlan
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1169-2018
- AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255
Terminated Class II Medical devices
- Recall started
- 18 January 2018
- Terminated
- 27 September 2023
- Reason
- A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.
- Distribution
- Worldwide distribution. US nationwide, Algeria, Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvi
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1151-2018
- AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
Terminated Class I Medical devices
- Recall started
- 12 January 2018
- Terminated
- 30 April 2024
- Reason
- A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
- Distribution
- Worldwide distribution. US Nationwide including Puerto Rico. OUS: FC 500: Japan, South Korea, China, Italy, Spain, United Kingdom, Germany, Bangladesh, Czech Republic, France, Pakistan, India, South Africa, Austria, Saudi Arabia, Algeria, Turkey, Belgium, Netherlands, Bulgaria, Sweden, Austral
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0471-2019, Z-0472-2019
- 1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics
- COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software
Terminated Class II Medical devices
- Recall started
- 4 December 2017
- Terminated
- 30 May 2023
- Reason
- A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous results for leukocytes.
- Distribution
- Worldwide Distribution and US Nationwide Distribution
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1893-2018
- iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to automate the urine chemistry analysis profile using iChem
Terminated Class II Medical devices
- Recall started
- 10 November 2017
- Terminated
- 21 December 2022
- Reason
- iQ200 Series Urine Microscopy Analyzer may intermittently fail to read urine sample dilution barcode labels causing erroneous results that could delay treatment for health conditions such as hematuria.
- Distribution
- United States: nationwide, including Puerto Rico; and Foreign: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Bermuda, Bolivia, Brazil, Brunei Darussalam, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, Guatemala, Hong Kon
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1366-2018
- iQ200 Series Urine Microscopy Analyzer with Barcode Reader Model NFT-2100, Catalog #s: 800-3046, 800-3047, 800-3052, 800-3053, 800-3802, 800-3803, 800-3900, 800-3920, 800-3925, 800-3931, 800-3933, 800-3934, 800-3935, 800
Terminated Class III Medical devices
- Recall started
- 6 November 2017
- Terminated
- 4 November 2019
- Reason
- Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
- Distribution
- Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0308-2018, Z-0310-2018, Z-0309-2018
- Access 2 Immunoassay System, Catalog Number 81600N. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
- UniCel DxC 600i SYNCHRON Access Clinical Systems; A25656, A25638. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human
- Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in huma
Terminated Class II Medical devices
- Recall started
- 24 August 2017
- Terminated
- 28 August 2020
- Reason
- Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used.
- Distribution
- Worldwide Distribution in the states of AL, AK, AZ, AK, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IO, KS, KY, LA, MA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NM, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VI, WA, WV, WI, and WY and the countries of Brazil, Canada, Costa Rica, El Salvador, Guam,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0605-2018
- Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman Coulter AU series of chemistry analyzers. Beckman Coulte
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.