Bayer Medical Care, Inc. recalls

8 recalls recorded since 2018, most recently on 31 July 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)3
Terminated or completed5
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
31 July 2026
Reason
Some devices may not have been properly sterilized due to a manufacturing sterilization issue.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-3147-2026

Ongoing Class II Medical devices

Recall started
24 April 2026
Reason
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2296-2026

Ongoing Class II Medical devices

Recall started
30 November 2022
Reason
Black specks may be present on the packaged needles
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, IN, MN, NY, OH, TX, VA and the country of China.
Initiated
Voluntary: Firm initiated
Recall number
Z-0910-2023

Terminated Class II Medical devices

Recall started
14 September 2021
Terminated
19 January 2023
Reason
Labeling error: MEDRAD Mark 7 Arterion Injection System Syringes (ART 700 SYR) labeled as Twist & Go Syringes (TAG 150 SYR)
Distribution
Worldwide distribution - US Nationwide and the countries of AU, BY, DE, DK, GB, HR, NL, NZ, RU, SE, TH.
Initiated
Voluntary: Firm initiated
Recall number
Z-0199-2022

Terminated Class II Medical devices

Recall started
5 October 2020
Terminated
19 March 2021
Reason
Specific lots of MEDRAD Stellant FLEX 150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.
Distribution
Distributed to healthcare facilities nationwide throughout the US.
Initiated
Voluntary: Firm initiated
Recall number
Z-0483-2021

Terminated Class II Medical devices

Recall started
14 December 2018
Terminated
16 June 2020
Reason
If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.
Distribution
The products were distributed US nationwide. The products were distributed to the following foreign countries: Algeria, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Cote D'Ivoire, Cuba, Curacao, Cyprus, Czech Republic, Denma
Initiated
Voluntary: Firm initiated
Recall number
Z-0745-2019

Terminated Class II Medical devices

Recall started
7 September 2018
Terminated
11 July 2019
Reason
Bayer has received reports of loose particulate in the filter when the dust cap is removed. Bayer conducted further testing confirming the dust cap may generate particulate which may potentially enter the fluid path.
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, and WI. Foreign countries: UAE, Austria, Australia, Bosnia and Herzegovina, Belgium, Bahra
Initiated
Voluntary: Firm initiated
Recall number
Z-0097-2019

Terminated Class II Medical devices

Recall started
28 June 2018
Terminated
26 November 2018
Reason
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Distribution
The devices were distributed to the following US states: OH, TX, and WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2961-2018, Z-2962-2018, Z-2960-2018