Baxter Healthcare Corporation recalls

182 recalls recorded since 2018, most recently on 14 September 2026. 114 are listed as ongoing. Status and classification are as published by the FDA.

Recalls182
Ongoing (FDA)114
Terminated or completed68
Class I (FDA)34

Ongoing Class II Medical devices

Recall started
14 September 2026
Reason
Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi,
Initiated
Voluntary: Firm initiated
Recall number
Z-0151-2027

Ongoing Class II Medical devices

Recall started
18 August 2026
Reason
There is an error in the product description on the Unique Device Identification (UDI) label. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect product identification or selection, potentially resulting in use with an incompatible bed frame.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3264-2026, Z-3262-2026, Z-3263-2026

Ongoing Class I Drugs

Recall started
17 August 2026
Reason
Presence of particulate matter: Particulates identified as fiberglass
Distribution
Natonwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0849-2026

Ongoing Class I Drugs

Recall started
17 August 2026
Reason
Presence of Particulate Matter: particulate matter identified as stainless steel particles in the solution.
Distribution
U.S. Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
D-0850-2026

Ongoing Class II Medical devices

Recall started
11 August 2026
Reason
A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3065-2026, Z-3066-2026

Ongoing Class II Drugs

Recall started
6 August 2026
Reason
CGMP Deviations
Distribution
US Nationwide; Puerto Rico; and Chile.
Initiated
Voluntary: Firm initiated
Recall numbers
D-0815-2026, D-0811-2026, D-0821-2026, D-0813-2026, D-0812-2026, D-0807-2026, D-0814-2026, D-0808-2026, D-0810-2026, D-0817-2026, D-0806-2026, D-0809-2026 and 5 more

Ongoing Class II Medical devices

Recall started
3 August 2026
Reason
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
Distribution
US Nationwide distribution in the states of Colorado and Florida.
Initiated
Voluntary: Firm initiated
Recall number
Z-3257-2026

Ongoing Class I Drugs

Recall started
24 July 2026
Reason
Presence of Particulate Matter
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
D-0787-2026

Ongoing Class I Medical devices

Recall started
30 June 2026
Reason
Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
Distribution
US Nationwide. No OUS consignees.
Initiated
Voluntary: Firm initiated
Recall number
Z-2766-2026

Ongoing Class I Medical devices

Recall started
21 May 2026
Reason
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2550-2026, Z-2551-2026, Z-2548-2026, Z-2549-2026

Ongoing Class II Medical devices

Recall started
22 April 2026
Reason
Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Caroli
Initiated
Voluntary: Firm initiated
Recall number
Z-2262-2026

Ongoing Class II Medical devices

Recall started
21 April 2026
Reason
The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota, Maryland, and Pennsylvania and the countries of Canada, Hong Kong, India, Japan, Korea, Philippines, Taiwan, Thailand, Vietnam, Austria, Belgium, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iraq, Isra
Initiated
Voluntary: Firm initiated
Recall number
Z-2638-2026

Ongoing Class I Medical devices

Recall started
18 February 2026
Reason
Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
Distribution
Florida
Initiated
Voluntary: Firm initiated
Recall number
Z-1637-2026

Ongoing Class II Medical devices

Recall started
18 February 2026
Reason
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
Initiated
Voluntary: Firm initiated
Recall number
Z-1573-2026

Ongoing Class II Medical devices

Recall started
13 January 2026
Reason
Pumps were released without full testing being performed, including occlusion alarm testing.
Distribution
States: Ohio Countries: United States
Initiated
Voluntary: Firm initiated
Recall number
Z-1368-2026

Ongoing Class II Drugs

Recall started
22 December 2025
Reason
Presence of particulate matter: plastic particles from the bottle rim were observed floating into the solution
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-0296-2026

Ongoing Class II Medical devices

Recall started
28 November 2025
Reason
Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1121-2026

Ongoing Class II Medical devices

Recall started
28 October 2025
Reason
Replacement bladder kit for the Welch Allyn two-piece reusable blood pressure cuff may contain a Large Adult size bladder rather than a Thigh size bladder in the replacement kits.
Distribution
US Nationwide distribution in the states of California, Colorado, Ohio, and Utah.
Initiated
Voluntary: Firm initiated
Recall number
Z-0847-2026

Ongoing Class II Medical devices

Recall started
17 September 2025
Reason
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Distribution
Distribution US nationwide, Canada and Colombia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0233-2026

Ongoing Class II Medical devices

Recall started
29 August 2025
Reason
IV sets may leak.
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0190-2026, Z-0194-2026, Z-0195-2026, Z-0198-2026, Z-0193-2026, Z-0202-2026, Z-0204-2026, Z-0203-2026, Z-0189-2026, Z-0197-2026, Z-0200-2026, Z-0192-2026 and 4 more

Ongoing Class II Drugs

Recall started
20 August 2025
Reason
Discoloration
Distribution
U.S.A. Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-0596-2025

Ongoing Class II Medical devices

Recall started
8 August 2025
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2594-2025

Ongoing Class I Medical devices

Recall started
4 August 2025
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
Distribution
US distribution including Puerto Rico and OUS (International) Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2333-2025, Z-2332-2025, Z-2334-2025

Ongoing Class I Medical devices

Recall started
14 July 2025
Reason
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2173-2025

Ongoing Class II Medical devices

Recall started
14 July 2025
Reason
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Distribution
US distribution to states of: AZ, FL
Initiated
Voluntary: Firm initiated
Recall number
Z-2415-2025

Ongoing Class I Medical devices

Recall started
20 June 2025
Reason
Certain Spectrum infusion pumps may have an incorrect version of software.
Distribution
US States: CA, FL, IL, KS, MI, MS, NY, OH, OK, and VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2095-2025, Z-2096-2025

Ongoing Class I Medical devices

Recall started
30 May 2025
Reason
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2032-2025, Z-2031-2025

Ongoing Class II Medical devices

Recall started
12 May 2025
Reason
Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1964-2025

Ongoing Class II Medical devices

Recall started
9 May 2025
Reason
The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1962-2025

Ongoing Class II Medical devices

Recall started
6 May 2025
Reason
There is a potential for improper crimps on the signal wires of the actuators which could result in a loss of ability to lower the bed or the foot section of the bed.
Distribution
Worldwide distribution - US Nationwide and the countries of BAHRAIN, Canada, Chile, FRANCE, KUWAIT, NETHERLANDS, NORWAY, UNITED ARAB EMIRATES.
Initiated
Voluntary: Firm initiated
Recall number
Z-1816-2025

Ongoing Class I Medical devices

Recall started
24 April 2025
Reason
There is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered off with the set loaded.
Distribution
US and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1767-2025

Ongoing Class I Medical devices

Recall started
7 April 2025
Reason
A cybersecurity vulnerability was discovered through internal testing.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1726-2025, Z-1725-2025

Ongoing Class II Medical devices

Recall started
24 March 2025
Reason
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1558-2025, Z-1559-2025, Z-1560-2025

Ongoing Class II Medical devices

Recall started
21 March 2025
Reason
Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1564-2025, Z-1563-2025, Z-1567-2025, Z-1568-2025, Z-1565-2025, Z-1566-2025, Z-1562-2025

Ongoing Class II Medical devices

Recall started
3 March 2025
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Novum IQ syringe pumps listed below due to an incorrectly installed gasket. A gasket failure could lead to potential fluid ingress due to cleaning, or as a result of an IV fluid spill.
Distribution
US Nationwide distribution in the states of AR, CA, FL, IA, IN, LA, MO, MT, NC, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1423-2025

Ongoing Class II Medical devices

Recall started
25 February 2025
Reason
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1419-2025, Z-1413-2025, Z-1422-2025, Z-1410-2025, Z-1416-2025, Z-1418-2025, Z-1414-2025, Z-1421-2025, Z-1409-2025, Z-1417-2025, Z-1420-2025, Z-1411-2025 and 2 more

Ongoing Class II Medical devices

Recall started
25 February 2025
Reason
The air bladders inside the mattress may move out of position when the head of the bed is elevated, causing a dip in the mattress.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1468-2025

Ongoing Class I Medical devices

Recall started
5 February 2025
Reason
There is the potential for missing motor mounting screws, which may have occurred during the servicing process.
Distribution
US: Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1253-2025, Z-1254-2025

Ongoing Class I Medical devices

Recall started
20 December 2024
Reason
There is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Distribution
US, Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0884-2025, Z-0887-2025, Z-0886-2025, Z-0885-2025

Ongoing Class II Medical devices

Recall started
20 December 2024
Reason
Specific lots of Solution Sets with Duo-Vent Spike have the potential for inverted side clamps causing the tubing to be unusable.
Distribution
US Nationwide distribution in the states of MN, MO, VA, TX, CA, MA, FL, SC.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1011-2025, Z-1009-2025, Z-1010-2025

Ongoing Class II Medical devices

Recall started
19 November 2024
Reason
Devices were identified as released after repair without full testing being performed, which includes flow testing.
Distribution
US Nationwide distribution in the states of AL, IN.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0776-2025, Z-0775-2025

Ongoing Class II Drugs

Recall started
30 October 2024
Reason
Labeling: Missing Label
Distribution
Nationwide within the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0039-2025

Ongoing Class II Medical devices

Recall started
23 October 2024
Reason
There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0560-2025, Z-0559-2025, Z-0569-2025, Z-0562-2025, Z-0568-2025, Z-0564-2025, Z-0561-2025, Z-0565-2025, Z-0563-2025, Z-0566-2025, Z-0567-2025

Ongoing Class I Medical devices

Recall started
21 October 2024
Reason
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
Distribution
Worldwide Distribution (US Nationwide distribution)
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0373-2025, Z-0375-2025, Z-0377-2025, Z-0379-2025, Z-0378-2025, Z-0382-2025, Z-0376-2025, Z-0381-2025, Z-0374-2025, Z-0380-2025

Ongoing Class I Medical devices

Recall started
16 October 2024
Reason
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
Distribution
US Nationwide distribution and worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0327-2025

Ongoing Class II Medical devices

Recall started
11 October 2024
Reason
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0323-2025, Z-0321-2025, Z-0324-2025, Z-0322-2025, Z-0320-2025, Z-0325-2025

Ongoing Class II Medical devices

Recall started
2 October 2024
Reason
The door on the Spectrum IQ Infusion pump may not be able to fully close.
Distribution
GA, ID
Initiated
Voluntary: Firm initiated
Recall number
Z-0348-2025

Ongoing Class II Medical devices

Recall started
19 September 2024
Reason
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0216-2025, Z-0217-2025, Z-0215-2025, Z-0214-2025, Z-0211-2025, Z-0213-2025, Z-0210-2025, Z-0212-2025, Z-0218-2025

Ongoing Class II Medical devices

Recall started
18 September 2024
Reason
The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.
Distribution
Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0228-2025

Ongoing Class I Medical devices

Recall started
12 September 2024
Reason
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
Distribution
US nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0046-2025

Ongoing Class I Medical devices

Recall started
20 August 2024
Reason
Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3085-2024, Z-3083-2024, Z-3084-2024, Z-3082-2024

Ongoing Class I Drugs

Recall started
1 August 2024
Reason
Microbial Contamination of Sterile Products; out of limit results obtained for endotoxin testing.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0649-2024

Ongoing Class II Medical devices

Recall started
17 July 2024
Reason
The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital s EMR system.
Distribution
US, Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2582-2024, Z-2581-2024

Ongoing Class II Medical devices

Recall started
28 June 2024
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2512-2024

Ongoing Class II Drugs

Recall started
17 June 2024
Reason
Lack of Assurance of Sterility: Potential presence of leaks originating from the Connector Assembly component.
Distribution
US Nationwide, Canada
Initiated
Voluntary: Firm initiated
Recall number
D-0587-2024

Ongoing Class II Medical devices

Recall started
14 June 2024
Reason
Improperly performed testing prior to release
Distribution
US Nationwide distribution in the state of Florida.
Initiated
Voluntary: Firm initiated
Recall number
Z-2530-2024

Ongoing Class II Medical devices

Recall started
13 June 2024
Reason
The Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU).
Distribution
Distributed to United States, Canada, Australia, Malaysia, New Zealand, Philippines, Austria, Belgium, France, Germany, Italy, Netherlands, Norway, Spain, Sweden, and United Kingdom. UPDATED PER EMAIL DATED 7/24/2024: United States (including Puerto Rico and Guam), Canada, Australia, Brunei, Hong
Initiated
Voluntary: Firm initiated
Recall number
Z-2339-2024

Ongoing Class II Medical devices

Recall started
4 June 2024
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2314-2024, Z-2315-2024, Z-2313-2024

Ongoing Class II Medical devices

Recall started
3 June 2024
Reason
During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
Distribution
Distribution to Illinois.
Initiated
Voluntary: Firm initiated
Recall number
Z-2323-2024

Ongoing Class I Medical devices

Recall started
30 May 2024
Reason
Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
Distribution
Nationwide distribution. International distribution to Canada, France, and Germany.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2321-2024, Z-2322-2024

Ongoing Class I Medical devices

Recall started
29 May 2024
Reason
Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2140-2024

Ongoing Class II Medical devices

Recall started
8 May 2024
Reason
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Distribution
Distribution was made nationwide. There was also government/military distribution. Foreign distribution was made to Australia, Canada, China, Japan, Qatar, Saudi Arabia, Singapore, Taiwan, Turkey, and United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall number
Z-2151-2024

Ongoing Class II Medical devices

Recall started
9 April 2024
Reason
There is a potential for cracks on the mount of the front panel of the device.
Distribution
US Nationwide distribution in the states of Ohio, Wisconsin.
Initiated
Voluntary: Firm initiated
Recall number
Z-1810-2024

Ongoing Class II Medical devices

Recall started
26 March 2024
Reason
Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
Distribution
Global, US: NY, IL
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1509-2024, Z-1511-2024, Z-1510-2024

Ongoing Class II Medical devices

Recall started
20 March 2024
Reason
One device was improperly performed testing prior to release from a Service Center.
Distribution
US Nationwide distribution in the state of Massachusetts.
Initiated
Voluntary: Firm initiated
Recall number
Z-1775-2024

Ongoing Class II Medical devices

Recall started
15 March 2024
Reason
Product was distributed in the United States without proper regulatory approval.
Distribution
US Nationwide distribution in the states of LA, GA, IN, VA FL, NY, CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1689-2024

Ongoing Class II Medical devices

Recall started
11 March 2024
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
Distribution
US Nationwide distribution in the state of Florida.
Initiated
Voluntary: Firm initiated
Recall number
Z-1329-2024

Ongoing Class II Medical devices

Recall started
5 March 2024
Reason
Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as potentially impacted by these service inconsistencies and therefore, Baxter will need to reassess the impacted mattresses to confirm they have been corrected as identified by the service order, and to confirm the correct serial numbers have been identified.
Distribution
US Nationwide distribution in the state of Colorado.
Initiated
Voluntary: Firm initiated
Recall number
Z-1400-2024

Ongoing Class II Medical devices

Recall started
28 February 2024
Reason
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Distribution
US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1378-2024, Z-1379-2024, Z-1377-2024, Z-1380-2024

Ongoing Class II Medical devices

Recall started
22 February 2024
Reason
The Surgical Tools and Cardiovascular Specialty marketing brochures distributed for Flo-Rester Disposable Internal Vessel Occluder, Flo-Thru Intraluminal Shunt, and Vascular Probe Intravascular Probe contain inaccurate content that does not align with the products Instructions for Use (IFU).
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Ireland, United Kingdom, and Hong Kong.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1522-2024, Z-1523-2024, Z-1524-2024

Ongoing Class II Medical devices

Recall started
14 February 2024
Reason
Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient harm and a delay in surgery.
Distribution
US Nationwide distribution in the state of Alabama.
Initiated
Voluntary: Firm initiated
Recall number
Z-1289-2024

Ongoing Class II Medical devices

Recall started
8 February 2024
Reason
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Clearlink Non-DEHP solution set listed below due to potential leaks originating from the drip chamber.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1291-2024

Ongoing Class II Drugs

Recall started
23 January 2024
Reason
Incorrect product concentration on the overwrap label: The overwrap label incorrectly identified the product strength as 4 mg / 250 mL; however, the primary bag label correctly identified the product strength as 8 mg / 250 mL.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0336-2024

Ongoing Class I Medical devices

Recall started
22 December 2023
Reason
An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volume.
Distribution
US: AL, AZ, CA, FL, GA, IA, KY, MD, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OR, SD, TX, WA, WI,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1001-2024, Z-1002-2024

Ongoing Class II Medical devices

Recall started
8 December 2023
Reason
Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Distribution
United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0707-2024, Z-0705-2024, Z-0706-2024

Ongoing Class II Medical devices

Recall started
6 December 2023
Reason
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0663-2024, Z-0661-2024, Z-0666-2024, Z-0667-2024, Z-0656-2024, Z-0668-2024, Z-0657-2024, Z-0669-2024, Z-0660-2024, Z-0658-2024, Z-0659-2024, Z-0664-2024 and 2 more

Ongoing Class II Medical devices

Recall started
15 November 2023
Reason
There is to an increase in complaints related to failure alarms for Wet Cassette Integrity Test occurring on the Amia devices.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0550-2024

Completed Class II Drugs

Recall started
14 November 2023
Reason
Failed pH Specifications
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0146-2024

Ongoing Class I Medical devices

Recall started
13 October 2023
Reason
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
Distribution
US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-0151-2024

Ongoing Class II Medical devices

Recall started
4 October 2023
Reason
Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.
Distribution
US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA
Initiated
Voluntary: Firm initiated
Recall number
Z-0212-2024

Terminated Class II Drugs

Recall started
24 August 2023
Terminated
22 December 2025
Reason
CGMP Deviations: Product was exposed to temperatures exceeding the labeled storage conditions during transportation were released by mistake.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0048-2024

Completed Class III Drugs

Recall started
14 August 2023
Reason
Labeling: Missing Label; customer complaint received that labels were partially or completely peeled off injection vial.
Distribution
Nationwide in the U.S.A.
Initiated
Voluntary: Firm initiated
Recall number
D-1103-2023

Terminated Class II Drugs

Recall started
11 August 2023
Terminated
16 July 2024
Reason
Lack of Assurance of Sterility: Potential presence of leaks originating from the Luer component.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall numbers
D-1141-2023, D-1140-2023

Ongoing Class II Medical devices

Recall started
10 August 2023
Reason
Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AL, AR, AZ, CA, CA, CT, FL, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, MO, MS, MT, NE, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Canada and Bermuda.
Initiated
Voluntary: Firm initiated
Recall number
Z-2647-2023

Ongoing Class II Medical devices

Recall started
10 August 2023
Reason
Product is being recalled due to the improper placement of a copper tape on the back of the liquid-crystal display.
Distribution
US Nationwide distribution in the state of CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2581-2023

Ongoing Class II Medical devices

Recall started
19 June 2023
Reason
An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall number
Z-2208-2023

Ongoing Class I Medical devices

Recall started
15 June 2023
Reason
A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2104-2023, Z-2103-2023

Ongoing Class II Medical devices

Recall started
19 May 2023
Reason
Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.
Distribution
US Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2007-2023, Z-2006-2023

Ongoing Class II Medical devices

Recall started
18 May 2023
Reason
Potential for the device to crack where the Chest Base Prone support attaches to the carbon fiber operating room (OR) table rail.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, China, Hong Kong, India, Japan, Korea, Malaysia, Singapore, Taiwan, Thailand, Belgium, Croatia, Denmark, Egypt, France, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Netherlands, Norway, Poland, Qatar, Saudi Arabia, So
Initiated
Voluntary: Firm initiated
Recall number
Z-2109-2023

Ongoing Class II Medical devices

Recall started
31 March 2023
Reason
There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Qatar.
Initiated
Voluntary: Firm initiated
Recall number
Z-1482-2023

Ongoing Class II Medical devices

Recall started
27 March 2023
Reason
The product is brittle and potentially crumbling upon handling or when being removed from its package.
Distribution
Worldwide distribution - US Nationwide and the countries of China, Germany, Sweden, Norway, Netherlands, Greece, Switzerland, France
Initiated
Voluntary: Firm initiated
Recall number
Z-1461-2023

Ongoing Class II Medical devices

Recall started
20 March 2023
Reason
ExataMed 20 mL oral dispenser component of the Clear Pharmacy Pack may have particulate matter.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall number
Z-1420-2023

Completed Class II Drugs

Recall started
15 March 2023
Reason
Lack of Assurance of Sterility: Potential presence of leaks originating from the Luer component.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall numbers
D-0543-2023, D-0542-2023

Ongoing Class II Medical devices

Recall started
22 February 2023
Reason
The electrical safety testing was not properly performed on the impacted devices and additional testing is required
Distribution
US distribution to AL, AR, AZ, FL, IL, ND, NY, OH, TX, WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1316-2023

Ongoing Class II Medical devices

Recall started
1 February 2023
Reason
MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.
Distribution
US Nationwide distribution and Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1213-2023

Ongoing Class I Medical devices

Recall started
25 January 2023
Reason
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1141-2023, Z-1140-2023, Z-1139-2023, Z-1142-2023

Ongoing Class II Medical devices

Recall started
23 November 2022
Reason
Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0762-2023, Z-0763-2023, Z-0764-2023

Ongoing Class II Medical devices

Recall started
23 November 2022
Reason
Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer
Distribution
Worldwide distribution. US Nationwide including Guam and St. Thomas USVI; Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecua
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0761-2023, Z-0760-2023

Terminated Class II Medical devices

Recall started
31 October 2022
Terminated
6 March 2025
Reason
One ELI 380 Electrocardiograph intended to be used for a pilot program was shipped to a customer by mistake on 26 October 2022.
Distribution
TX
Initiated
Voluntary: Firm initiated
Recall number
Z-1013-2023

Ongoing Class II Medical devices

Recall started
18 October 2022
Reason
The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, HI, IN, LA, MA, MO, NC, NM, NV, NY, OK, OR, PA, SC. TN, TX, WA, and WI. (UPDATED: States of PA and SC and there was government distribution.) There was no military distribution. Foreign distribution was made to Canada, Qatar
Initiated
Voluntary: Firm initiated
Recall number
Z-0231-2023

Ongoing Class I Medical devices

Recall started
30 September 2022
Reason
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Distribution
US
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0143-2023, Z-0142-2023, Z-0141-2023, Z-0145-2023, Z-0144-2023

Terminated Class II Medical devices

Recall started
20 September 2022
Terminated
4 August 2026
Reason
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
Distribution
MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0180-2023, Z-0177-2023, Z-0175-2023, Z-0179-2023, Z-0176-2023, Z-0178-2023

Ongoing Class II Medical devices

Recall started
14 September 2022
Reason
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0085-2023, Z-0084-2023, Z-0081-2023, Z-0082-2023, Z-0083-2023

Ongoing Class III Medical devices

Recall started
2 September 2022
Reason
Labeling has incorrect or incomplete Unique Device Identification (UDI) information; and the barcodes are not compliant with GS-1 standards.
Distribution
US Nationwide Distribution and government/military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1817-2022

Terminated Class I Medical devices

Recall started
9 August 2022
Terminated
9 August 2024
Reason
Firm noted an increase in customer reports of leaks.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-1653-2022

Terminated Class II Drugs

Recall started
4 August 2022
Terminated
16 July 2024
Reason
Temperature Abuse: Product distributed in refrigerated trucks with labels attached to pallets indicating "Refrigerate Upon Arrival", however product is labeled to be stored in a controlled room temperature environment.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-1348-2022

Ongoing Class II Medical devices

Recall started
22 July 2022
Reason
There is a potential of the packaging not maintaining a sterile barrier for the tip caps.
Distribution
Worldwide distribution: US (nationwide) distribution and OUS (foreign) : Australia, Finland, Germany, Spain and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1578-2022, Z-1579-2022, Z-1576-2022, Z-1581-2022, Z-1577-2022, Z-1575-2022, Z-1580-2022

Ongoing Class II Medical devices

Recall started
14 July 2022
Reason
There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets.
Distribution
Worldwide distribution: US Nationwide, Canada, Uruguay, Argentina, Brazil, Colombia, Dominican Republic, Panama, Puerto Rico, Hong Kong, Taiwan, Korea, United Kingdom, Austria, Belgium, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Ireland, Israel, Italy, Kuwait, Netherlands, Oman, Pola
Initiated
Voluntary: Firm initiated
Recall number
Z-1570-2022

Ongoing Class II Medical devices

Recall started
24 June 2022
Reason
The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. There was government distribution but no military distribution. The country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1411-2022, Z-1413-2022, Z-1417-2022, Z-1416-2022, Z-1415-2022, Z-1414-2022, Z-1412-2022

Terminated Class I Medical devices

Recall started
22 June 2022
Terminated
9 October 2024
Reason
There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
Distribution
United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA
Initiated
Voluntary: Firm initiated
Recall number
Z-1376-2022

Ongoing Class II Medical devices

Recall started
31 May 2022
Reason
An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Distribution
nationwide and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1306-2022

Terminated Class II Medical devices

Recall started
17 May 2022
Terminated
4 May 2026
Reason
Packaging error.
Distribution
Worldwide distribution - US Nationwide and the countries of Japan, Canada, Australia, New Zealand, Spain, Brunei Darussalam, Brazil, India, Colombia, Czechia, Egypt, Turkey, South Africa, Argentina, Mexico, Congo, United Arab Emirates, El Salvador, South Korea, Germany, Cameroon, Guyana, Austria, B
Initiated
Voluntary: Firm initiated
Recall number
Z-1288-2022

Terminated Class I Medical devices

Recall started
26 April 2022
Terminated
20 August 2024
Reason
There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.
Distribution
US Nationwide distribution in the states of TX, FL, VA, KY, CA, AZ, WI, IL, NC, OH, MI, MS, MO, MN, AL, TN, SD, OK, NV, NJ, IN, CO, GA, WA, PA, IA, NE, NM, SC, ND, KS, WV, OR.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1139-2022, Z-1138-2022

Terminated Class II Medical devices

Recall started
21 April 2022
Terminated
27 May 2026
Reason
Certain Peritoneal Dialysis product codes do not have FDA regulatory clearance and were distributed within the United States in error.
Distribution
US Nationwide distribution in the states of CA, MO, MA, AZ, TX, FL, NY, MD, OH.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1145-2022, Z-1141-2022, Z-1146-2022, Z-1142-2022, Z-1143-2022, Z-1144-2022

Terminated Class II Medical devices

Recall started
13 April 2022
Terminated
20 December 2023
Reason
Firm discovered a firmware memory leak with a supplier-manufactured component.
Distribution
US Distribution to states of: NC and VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1094-2022, Z-1093-2022

Terminated Class II Medical devices

Recall started
1 April 2022
Terminated
10 July 2026
Reason
There is a potential for foreign matter.
Distribution
Distribution in the United States including AL, CA, KY, NY, OR, PA, and TX OUS distribution to Japan, Italy, Netherlands, Switzerland, and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-1019-2022

Ongoing Class II Medical devices

Recall started
15 March 2022
Reason
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
Distribution
Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0993-2022, Z-0992-2022, Z-0991-2022

Terminated Class II Medical devices

Recall started
25 February 2022
Terminated
15 May 2026
Reason
Devices distributed lacked regulatory clearance.
Distribution
US Nationwide distribution in the states of California, Colorado, District of Columbia, Delaware, Florida, Georgia, Illinois, Louisiana, Michigan, Missouri, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, and Texas.
Initiated
Voluntary: Firm initiated
Recall number
Z-0831-2022

Terminated Class II Medical devices

Recall started
25 February 2022
Terminated
2 March 2026
Reason
Product is mislabeled as self-righting Luer slip tip caps, yellow.
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, GA, IL, IN, MO, PA, TX, and VA. The countries of Canada, Austria, Germany, Netherlands, Qatar, Saudi Arabia, and Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-0948-2022

Ongoing Class II Medical devices

Recall started
21 February 2022
Reason
An incorrect center bolt was used in the assembly of the product which could result in a patient fall during use.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, ND, NE, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. There was government distribution but no military distribution. The countries of Austria,
Initiated
Voluntary: Firm initiated
Recall number
Z-0829-2022

Terminated Class II Medical devices

Recall started
14 February 2022
Terminated
2 July 2026
Reason
Potential damage to the power cord may cause the power conductors to be exposed, thereby creating a risk of shock from the power mains.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Cameroon, Canada, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Jamaica, Jordan, Kuwait, Mexico, Netherlands, New Zealand, Norway, Oman, P
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0821-2022, Z-0820-2022

Ongoing Class II Medical devices

Recall started
7 January 2022
Reason
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0576-2022, Z-0572-2022, Z-0570-2022, Z-0574-2022, Z-0578-2022, Z-0580-2022, Z-0579-2022, Z-0573-2022, Z-0577-2022, Z-0571-2022, Z-0575-2022

Ongoing Class I Medical devices

Recall started
29 December 2021
Reason
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Distribution
Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0529-2022, Z-0530-2022

Ongoing Class II Medical devices

Recall started
29 September 2021
Reason
If the operator initiates therapy with a saved prescription profile and makes a change to the prescription after a disposable filter change using the Same Patient button, the system may suggest values from the original prescription profile, rather than the current prescription.
Distribution
Distributed nationwide to AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC,SD, TN, TX, UT, VA, WA, WI and internationally to Mexico, Brazil, Canada, Australia, Singapore, Hong Kong, Vietnam, Austria, Belgium, Denmark, Finl
Initiated
Voluntary: Firm initiated
Recall number
Z-0215-2022

Terminated Class II Medical devices

Recall started
22 July 2021
Terminated
17 January 2024
Reason
Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2379-2021, Z-2378-2021

Terminated Class I Medical devices

Recall started
7 July 2021
Terminated
20 June 2024
Reason
Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
Distribution
US Nationwide Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2165-2021

Terminated Class II Medical devices

Recall started
30 June 2021
Terminated
10 April 2024
Reason
Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.
Distribution
Worldwide distribution - US Nationwide distribution and the country of Japan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2173-2021, Z-2174-2021

Terminated Class II Medical devices

Recall started
15 June 2021
Terminated
12 January 2024
Reason
There is a discrepancy in the expiry date encoded into the 2D barcode. The correct expiration date is the human-readable date
Distribution
United States, Australia, Hong Kong, New Zealand, Belgium, Cyprus, Ireland, Italy, Netherlands, Norway, Portugal, Spain, Sweden, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2051-2021

Terminated Class I Medical devices

Recall started
4 June 2021
Terminated
25 September 2024
Reason
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Distribution
United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
Initiated
Voluntary: Firm initiated
Recall number
Z-2110-2021

Terminated Class II Medical devices

Recall started
14 April 2021
Terminated
10 May 2024
Reason
There is a potential leak between the venous patient connector and patient's hemodialysis access.
Distribution
Worldwide distribution - United States distribution and the countries of Canada, Korea, and Norway.
Initiated
Voluntary: Firm initiated
Recall number
Z-1684-2021

Terminated Class II Medical devices

Recall started
19 March 2021
Terminated
15 July 2022
Reason
Potentially defective component in the AC power adapter which may fail over time.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CT, FL, IA, IN, LA, MA, MD, ME, MN, MO, MS, MT, NC, NJ, NY, OH, PA, RI, SC, TN, TX and the country of Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-1487-2021

Terminated Class II Medical devices

Recall started
28 January 2021
Terminated
13 June 2024
Reason
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Distribution
Worldwide distribution - US Nationwide distribution including Puerto Rico and Guam and the countries of Albania, Algeria, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia Herzegovina, Bulgaria, Canada, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1278-2021, Z-1279-2021, Z-1280-2021, Z-1277-2021

Terminated Class II Medical devices

Recall started
7 January 2021
Terminated
30 April 2024
Reason
Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance. Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Fin
Initiated
Voluntary: Firm initiated
Recall number
Z-1151-2021

Terminated Class II Medical devices

Recall started
6 January 2021
Terminated
4 May 2026
Reason
Battery performance issues related to prolonged storage.
Distribution
Domestic Distribution: ST, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Puerto Rico. International Distribution: Canada, Barbados, Bermu
Initiated
Voluntary: Firm initiated
Recall number
Z-1150-2021

Ongoing Class II Medical devices

Recall started
30 December 2020
Reason
complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1121-2021, Z-1120-2021, Z-1118-2021, Z-1122-2021, Z-1123-2021, Z-1119-2021

Ongoing Class II Medical devices

Recall started
28 December 2020
Reason
Potential for the battery to lose its ability to be recharged.
Distribution
Distributed nationwide to AZ, FL, OH, NY, GA, VA, MA, WI, MO, MT, WV, IL, NC, NJ, AR, CA, CO, CT, IA, ID, IN, KS, LA, MD, MI, MN, MS, NE, NH, NM, NV, OK, OR, PA, RI, WY, WA, VT, TX, TN, SC, SD
Initiated
Voluntary: Firm initiated
Recall number
Z-0939-2021

Terminated Class II Medical devices

Recall started
21 December 2020
Terminated
11 July 2024
Reason
There were customer reports of separation between the tubing and Male Luer Lock Adaptor.
Distribution
Worldwide distribution - US Nationwide distribution in the states of SC, OH, MI, CA, TX, and the countries of UK, Ireland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1102-2021

Terminated Class II Medical devices

Recall started
11 December 2020
Terminated
24 March 2023
Reason
VERITAS Collagen Matrix 12x25 cm Patches may have undersized width dimensions.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, IL, MA, MS, OH, OR, TX, VA, WA, WV and the countries of Australia, Poland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0807-2021

Ongoing Class II Medical devices

Recall started
23 November 2020
Reason
There is a potential software error during programming.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Barbados, Jamaica, Bermuda, Guyana, Bahamas, and Trinidad and Tobago.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0712-2021, Z-0714-2021, Z-0715-2021, Z-0713-2021

Terminated Class II Medical devices

Recall started
10 November 2020
Terminated
24 March 2023
Reason
Potential for no-flow and leaks under the twist clamp.
Distribution
Worldwide and US (Nationwide), Canada, Brazil, Mexico, Bermuda, and Indonesia
Initiated
Voluntary: Firm initiated
Recall number
Z-0768-2021

Terminated Class II Drugs

Recall started
7 October 2020
Terminated
22 May 2023
Reason
Presence of particulate matter in solution - black and transparent particles
Distribution
Nationwide USA
Initiated
Voluntary: Firm initiated
Recall number
D-0076-2021

Terminated Class I Medical devices

Recall started
30 June 2020
Terminated
29 April 2024
Reason
Baxter is updating the device IFU for safety communication regarding cleaning practices.
Distribution
Distribution throughout the United States, St. Thomas USVI, and Puerto Rico. Foreign distribution including Canada, Caribbean (Barbados, Trinidad and Tobago, Jamaica, Bahamas, Bermuda, and Guyana)
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0121-2021, Z-0122-2021, Z-0123-2021, Z-0120-2021

Ongoing Class II Medical devices

Recall started
2 June 2020
Reason
A firmware issue may cause the GEM1020M-2 Flow Coupler Monitor to intermittently shut down when a WiFi connection is not established.
Distribution
U.S. distribution only: VA, GA, MD, PA, DC, WI, LA, MN, FL, IN, MI, AZ, TX, MS, MA, NY, SC, WI, WA, MI, CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2474-2020

Terminated Class II Medical devices

Recall started
17 April 2020
Terminated
7 February 2023
Reason
Potential disconnection of tubing set.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Countries of Australia, Austria, Bahrain, B
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2196-2020, Z-2193-2020, Z-2194-2020, Z-2195-2020

Terminated Class II Medical devices

Recall started
7 April 2020
Terminated
9 November 2021
Reason
Prismaflex product code 114870 does not have FDA regulatory clearance and was inadvertently distributed.
Distribution
Distribution only to Puerto Rico and Guam
Initiated
Voluntary: Firm initiated
Recall number
Z-2066-2020

Terminated Class II Medical devices

Recall started
5 March 2020
Terminated
2 August 2021
Reason
The Coupler size labeling of the outer tray may not match the Couper size labeling of the inner tray in the GEM Microvascular Coupler Device
Distribution
Worldwide distribution - United States Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, and WI and countries of Saudi Arabia, Italy, Austria, Argentina, Japan, Israel,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1686-2020, Z-1687-2020, Z-1685-2020

Terminated Class II Medical devices

Recall started
6 February 2020
Terminated
1 June 2021
Reason
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
Distribution
Distribution in US, Puerto Rico
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1830-2020, Z-1831-2020, Z-1832-2020, Z-1828-2020, Z-1829-2020, Z-1833-2020

Ongoing Class II Medical devices

Recall started
12 December 2019
Reason
Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
Distribution
Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0793-2020, Z-0794-2020

Terminated Class II Drugs

Recall started
19 November 2019
Terminated
4 February 2021
Reason
Lack of Assurance of Sterility: Product has the potential to leak.
Distribution
Nationwide within the United States
Initiated
Voluntary: Firm initiated
Recall number
D-0517-2020

Terminated Class II Medical devices

Recall started
15 November 2019
Terminated
20 November 2020
Reason
There is the potential presence of particular matter in the header caps of ten lots the Revaclear 300 Dialyzer.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1409-2020

Ongoing Class II Medical devices

Recall started
4 October 2019
Reason
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. International distribution to Australia, Bangladesh, Brunei, China,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0276-2020, Z-0277-2020, Z-0279-2020, Z-0280-2020, Z-0278-2020

Terminated Class II Medical devices

Recall started
19 September 2019
Terminated
11 December 2020
Reason
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
Distribution
Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand
Initiated
Voluntary: Firm initiated
Recall number
Z-0247-2020

Terminated Class II Medical devices

Recall started
18 September 2019
Terminated
16 February 2021
Reason
Potential ability to leak once used for compounding.
Distribution
Worldwide distribution - UD Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. The countries of Canada, Aust
Initiated
Voluntary: Firm initiated
Recall number
Z-0167-2020

Terminated Class III Medical devices

Recall started
13 August 2019
Terminated
23 November 2020
Reason
Exactamed 5 mL Oral Dispenser Pharmacy Packs may contain both 5 mL and 3 mL oral dispensers.
Distribution
Worldwide distribution - including US Nationwide distribution, and country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0050-2020

Terminated Class II Drugs

Recall started
18 July 2019
Terminated
19 April 2021
Reason
Lack of Assurance of Sterility: Bags have potential to leak.
Distribution
Nationwide in the United States and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
D-1514-2019

Terminated Class III Medical devices

Recall started
27 June 2019
Terminated
2 November 2020
Reason
EXACTAMED 5 mL Oral Dispenser Pharmacy Packs may contain both 5 mL and 10 mL oral dispensers.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2432-2019

Terminated Class II Drugs

Recall started
13 June 2019
Terminated
29 April 2020
Reason
Failed Stability Specifications
Distribution
Nationwide within the United States
Initiated
Voluntary: Firm initiated
Recall number
D-1435-2019

Terminated Class II Medical devices

Recall started
31 May 2019
Terminated
5 October 2020
Reason
There is a potential presence of ruptured dialyzer fibers which may lead to a blood leak during treatment.
Distribution
US Nationwide Distribution and Internationally to: Canada and Bermuda.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1909-2019, Z-1908-2019

Terminated Class II Drugs

Recall started
2 January 2019
Terminated
29 April 2020
Reason
Lack of Assurance of Sterility: Confirmed customer complaints for leaks on the tubing.
Distribution
Nationwide in the USA to Wholesaler/Distributors, Healthcare Facilities and Home Patients.
Initiated
Voluntary: Firm initiated
Recall number
D-0351-2019

Terminated Class II Medical devices

Recall started
11 December 2018
Terminated
12 February 2021
Reason
Potential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy
Distribution
Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Canad
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0850-2019, Z-0849-2019

Ongoing Class II Medical devices

Recall started
2 November 2018
Reason
Potential presence of rust on Coupler Forceps (GEM4183C).
Distribution
Nationwide distribution to CA, FL, GA, KS, LA, MA, MS, NE, NJ, NV, NY, OH, PA, TX, UT, VA, WA, WV. International distribution to United Kingdom, Germany, South Africa, Japan, and Australia
Initiated
Voluntary: Firm initiated
Recall number
Z-0680-2019

Terminated Class III Drugs

Recall started
15 October 2018
Terminated
12 August 2022
Reason
Failed pH Specifications: Upward shift in the pH of the solution within the shelf life of the impacted lots.
Distribution
Distributed nationwide in the USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-0222-2019

Terminated Class II Medical devices

Recall started
10 October 2018
Terminated
22 July 2020
Reason
Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.
Distribution
U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0030-2020

Terminated Class III Drugs

Recall started
25 September 2018
Terminated
26 June 2020
Reason
Superpotent Drug: High out of specification results for levofloxacin resulting in increased concentration of solution.
Distribution
Nationwide USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-1215-2018

Terminated Class II Medical devices

Recall started
18 September 2018
Terminated
19 August 2020
Reason
If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.
Distribution
NJ
Initiated
Voluntary: Firm initiated
Recall number
Z-0107-2019

Terminated Class II Medical devices

Recall started
12 September 2018
Terminated
2 November 2020
Reason
Potential ability to leak once used for compounding.
Distribution
Worldwide distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Countries of Canada, Australia, New Zeal
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0363-2019, Z-0367-2019, Z-0365-2019, Z-0366-2019, Z-0364-2019

Terminated Class II Drugs

Recall started
6 September 2018
Terminated
1 October 2020
Reason
cGMP Deviation
Distribution
Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
D-1190-2018

Terminated Class II Drugs

Recall started
21 August 2018
Terminated
30 October 2019
Reason
CGMP Deviations
Distribution
TX and MS only
Initiated
Voluntary: Firm initiated
Recall number
D-1128-2018

Terminated Class III Drugs

Recall started
1 August 2018
Terminated
16 July 2020
Reason
Superpotent Drug: High out of specification results for levofloxacin resulting in increased concentration of solution.
Distribution
Nationwide USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall numbers
D-1034-2018, D-1036-2018, D-1035-2018

Terminated Class II Medical devices

Recall started
12 July 2018
Terminated
31 March 2021
Reason
There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
Distribution
US and Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-2597-2018

Terminated Class II Medical devices

Recall started
21 June 2018
Terminated
23 May 2019
Reason
The flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate testing may not identify issues with pumps that could cause inaccuracies in flow rates which, in turn, could lead to over- or under- infusion. Potential risk to the patient depends on a number of factors including treatment and patient comorbidities. Depending on these factors, the patient could experience serious adverse health consequences.
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, FA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, VT, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1248-2019

Ongoing Class II Medical devices

Recall started
30 May 2018
Reason
Baxter r recalled this device due to the use of an incorrect plastic formulation in the manufacture of the blue luer component of the FLOSEAL malleable tips.
Distribution
Nationwide distribution to AK, AL, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Australia, Korea, Taiwan, Columbia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2957-2018

Terminated Class II Medical devices

Recall started
8 May 2018
Terminated
8 April 2019
Reason
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
Distribution
Worldwide Distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2113-2018, Z-2112-2018

Terminated Class II Medical devices

Recall started
25 April 2018
Terminated
16 December 2020
Reason
Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiration date,
Distribution
Worldwide Distribution - US Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1909-2018, Z-1908-2018, Z-1910-2018, Z-1911-2018

Terminated Class II Medical devices

Recall started
24 April 2018
Terminated
22 June 2021
Reason
Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.
Distribution
Worldwide distribution: United States (including Puerto Rico), and countries of: Canada, Australia, New Zealand, China, Bangladesh, Japan, Malaysia, Myanmar, Sri Lanka, India, Singapore, Vietnam, Hong Kong, Korea, Taiwan, Indonesia, Bruni, Philippines, Thailand, Argentina, Chile, Costa Rica, Dominic
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1926-2018, Z-1922-2018, Z-1925-2018, Z-1924-2018, Z-1923-2018

Terminated Class II Medical devices

Recall started
3 April 2018
Terminated
14 February 2020
Reason
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.
Distribution
US Nationwide & Puerto Rico
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1665-2018, Z-1667-2018, Z-1663-2018, Z-1664-2018, Z-1666-2018

Terminated Class II Medical devices

Recall started
6 February 2018
Terminated
2 April 2021
Reason
Affected lot of 5 mL oral dispensers was packaged in individual pouches that are mislabeled with the incorrect product code (H93876101) and volume (1mL).
Distribution
Distributed domestically to IL, MA, MD, OH, SD, TX, VA, WI. Distributed internationally to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0806-2018

Terminated Class II Medical devices

Recall started
21 December 2017
Terminated
4 August 2020
Reason
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
Distribution
US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.
Initiated
Voluntary: Firm initiated
Recall number
Z-0411-2018

Terminated Class I Drugs

Recall started
10 November 2017
Terminated
29 August 2018
Reason
Presence of Particulate Matter:Particulate identified as polyethylene, the primary constituent of the film and ports used to manufacture the bag in which product is packaged
Distribution
Nationwide within USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-0573-2018

Terminated Class II Medical devices

Recall started
6 October 2017
Terminated
11 April 2019
Reason
Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
Distribution
Worldwide Distribution - US, including the District of Columbia, Puerto Rico, and Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0971-2018

Terminated Class II Medical devices

Recall started
27 July 2017
Terminated
3 November 2020
Reason
A leak may allow for microbial contamination of the sterile fluid path.
Distribution
Distribution in the US and Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0485-2018

Terminated Class II Medical devices

Recall started
15 February 2017
Terminated
8 May 2019
Reason
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Algeria, Armenia, Australia, Bahrain, Belarus, Belgium, Canada, China, Cyprus, Denmark, Egypt, France, Germany, Great Britain, Greece, Ireland, Italy, Albania, Bulgaria, Croatia, Czech Republic, Estonia, Hong Kong, Hu
Initiated
Voluntary: Firm initiated
Recall number
Z-1280-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.