Avanos Medical, Inc. recalls

19 recalls recorded since 2018, most recently on 30 April 2026. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls19
Ongoing (FDA)6
Terminated or completed13
Class I (FDA)4

Ongoing Class II Medical devices

Recall started
30 April 2026
Reason
Lidocaine hydrochloride injection, the subject of a supplier recall for quality issues, were included into sterile kits and sets.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2337-2026, Z-2343-2026, Z-2338-2026, Z-2348-2026, Z-2346-2026, Z-2334-2026, Z-2344-2026, Z-2335-2026, Z-2339-2026, Z-2352-2026, Z-2347-2026, Z-2349-2026 and 8 more

Ongoing Class I Medical devices

Recall started
19 March 2025
Reason
Lack of sterility assurance for closed suction catheter systems
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, IL, IN,KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of South Korea, New Zealand, Canada, Mexico, Germany, England, Slovenia, France
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1528-2025, Z-1521-2025, Z-1523-2025, Z-1525-2025, Z-1527-2025, Z-1526-2025, Z-1524-2025, Z-1522-2025

Ongoing Class II Medical devices

Recall started
31 May 2024
Reason
The IFU was missing some risk information requested by another country to be contained in the IFU pertaining to potential complications.
Distribution
Worldwide distribution - US Nationwide - There was also government and military distribution and the countries of Canada and Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2422-2024

Ongoing Class III Medical devices

Recall started
4 March 2024
Reason
The expiration date was not printed on the outer shipping label, leading to expired syringes being shipped to customers.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, New Zealand, Canada, Switzerland, Netherlands, Singapore.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2037-2024, Z-2035-2024, Z-2036-2024

Ongoing Class I Medical devices

Recall started
16 January 2024
Reason
MIC* Gastric - Jejunal Feeding Tube Kits contain a sterile pre-filled syringe subsequently recalled by Nurse Assist.
Distribution
Worldwide distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including PR; and OUS (foreign) countries of: C
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1186-2024, Z-1187-2024

Ongoing Class II Medical devices

Recall started
13 October 2023
Reason
Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product did contain DEHP.
Distribution
US and South Africa
Initiated
Voluntary: Firm initiated
Recall number
Z-0360-2024

Completed Class II Medical devices

Recall started
1 August 2023
Reason
Avanos has received an increase in complaints from customers for COOLIEF RF Generators producing F100 and F101 faults.
Distribution
US Distribution to states of: AZ, CA, GA, IA, IN, KS, LA, MN, NC, NM, NY, OH, OR, TX, VA, VT, and OUS International distribution to country of: BRAZIL
Initiated
Voluntary: Firm initiated
Recall number
Z-2622-2023

Terminated Class II Medical devices

Recall started
24 April 2023
Terminated
2 August 2024
Reason
Some of the ambIT kits were potentially distributed without an air in-line filter.
Distribution
Worldwide distribution - US Nationwide distribution in the states of Texas and Oregon. The country of Australia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1684-2023, Z-1686-2023, Z-1688-2023, Z-1687-2023, Z-1685-2023

Terminated Class I Medical devices

Recall started
3 February 2023
Terminated
5 September 2024
Reason
Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
Distribution
US Nationwide distribution in the state of Oklahoma.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1190-2023, Z-1189-2023

Terminated Class II Medical devices

Recall started
3 February 2023
Terminated
19 August 2024
Reason
Incorrect expiration date.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Netherlands, France, Switzerland, Great Britain, Israel, and India.
Initiated
Voluntary: Firm initiated
Recall number
Z-1406-2023

Terminated Class II Medical devices

Recall started
14 September 2022
Terminated
31 August 2023
Reason
Sterile extension sets were distributed without an expiration date
Distribution
US: CA, CO, FL, MI, NM, OH, OR, TN, TX, UT, WI
Initiated
Voluntary: Firm initiated
Recall number
Z-0195-2023

Terminated Class I Medical devices

Recall started
21 March 2022
Terminated
9 September 2024
Reason
Product labeling is being modified. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol.
Distribution
Worldwide Distribution. US Nationwide, Australia, Austria, Canada, Germany, Ireland, Kuwait, New Zealand, India, Japan, Malaysia, Netherlands, Panama, Singapore, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1032-2022

Terminated Class II Medical devices

Recall started
22 December 2021
Terminated
15 August 2023
Reason
The External Retention Bolster contained in the kit may be a larger diameter than the tube.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Chile.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0566-2022, Z-0565-2022

Terminated Class II Medical devices

Recall started
22 November 2021
Terminated
31 January 2023
Reason
Product was distributed with incorrect expiration date.
Distribution
Worldwide distribution - US Nationwide distribution in the states of Oklahoma and Ohio; The countries of China, France, Germany, Great Britain, Greece, Italy, and Netherlands.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0846-2022, Z-0847-2022, Z-0845-2022

Terminated Class II Medical devices

Recall started
16 October 2020
Terminated
30 September 2022
Reason
Detachment of the Male AIV Adaptor from the Y-Connector.
Distribution
Domestic distribution only: CA, CO, IL, LA, OH, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0540-2021

Terminated Class II Medical devices

Recall started
29 July 2020
Terminated
31 January 2023
Reason
Devices had cracked or detached from the feeding tube body
Distribution
US states: OK IL WI PA TX IL PA NJ OK LA OUS: Belgium
Initiated
Voluntary: Firm initiated
Recall number
Z-2969-2020

Terminated Class II Medical devices

Recall started
16 March 2020
Terminated
31 August 2023
Reason
The CORFLO PEG tube may become blocked at the center of the adapter (occluded) potentially causing a procedure to delayed or aborted.
Distribution
Worldwide and US Nationwide Distribution: AZ, CA, CT, FL, GA, IA, ID, IL, LA, MO, MN, MD, NY, NM, NV, NJ, OR, OK, OH, RI, TN, TX, WA, WI, ***Foreign: Distributed to Australia, Belgium, Canada, France, Germany, Ireland, Italy, Netherlands, Romania, United Arab Emirates, United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-1822-2020

Terminated Class II Medical devices

Recall started
5 November 2019
Terminated
16 August 2023
Reason
Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.
Distribution
Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland
Initiated
Voluntary: Firm initiated
Recall number
Z-0788-2020

Terminated Class II Medical devices

Recall started
8 October 2019
Terminated
16 March 2022
Reason
The connector piece to the PEG tube does not pass over the guidewire.
Distribution
Distribution to CT, IN, MD, NJ, PA, VA. International distribution to Great Britain.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0371-2020, Z-0370-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.