Alfreton, United Kingdom
Autochair recalls
1 recall recorded since 2018, most recently on 7 March 2018. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 7 March 2018
- Reason
- Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.
- Distribution
- PA, and worldwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1420-2018
- Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.