Ascensia Diabetes Care US, Inc. recalls

1 recall recorded since 2018, most recently on 19 May 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
19 May 2023
Reason
Ascensia Diabetes Care has determined through customer complaints that 580 of 2900 meters from lot number DM01T033P are reconfigured from mg/dL to mmol/L. Among them, 579 meters distributed in the United States were packaged into the meter kit and had incorrect factory-set unit of measure where the meters display glucose results in mmol/L rather than mg/dL. The standard unit of measurement for the United States is mg/dL. If a consumer does not notice the incorrect unit of measurement, it is possible that the meters glucose measurement will be read as a lower blood glucose measurement than expected, and this may result in the patients glucose level remaining high, which can potentially lead to an injury requiring immediate medical intervention.
Distribution
Domestic: CA, CO, FL, IN, MO, NJ, NY, OH, RI, SC, TN, TX, & VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2211-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.