Recall Record

Lower Hutt, New Zealand

AROA Biosurgery, LTD. recalls

2 recalls recorded since 2018, most recently on 27 November 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Terminated Class II Medical devices

Recall started
27 November 2018
Terminated
9 April 2019
Reason
Potential for pouch seal failure
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0378-2019, Z-0377-2019, Z-0379-2019

Terminated Class II Medical devices

Recall started
6 April 2018
Terminated
15 April 2019
Reason
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
Distribution
The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0421-2019, Z-0416-2019, Z-0429-2019, Z-0423-2019, Z-0432-2019, Z-0418-2019, Z-0435-2019, Z-0431-2019, Z-0430-2019, Z-0427-2019, Z-0415-2019, Z-0428-2019 and 9 more

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.