Armstrong Medical Services Limited recalls

1 recall recorded since 2018, most recently on 5 August 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)1

Terminated Class I Medical devices

Recall started
5 August 2021
Terminated
17 September 2024
Reason
Some devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthetised patient, which could cause a delay in treatment. Not all defective canisters will be identified by the pre-use test.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Canada, Finland, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Portugal, United Kingdom.