Armstrong Medical Services Limited recalls
1 recall recorded since 2018, most recently on 5 August 2021. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)1
Terminated Class I Medical devices
- Recall started
- 5 August 2021
- Terminated
- 17 September 2024
- Reason
- Some devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthetised patient, which could cause a delay in treatment. Not all defective canisters will be identified by the pre-use test.
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Canada, Finland, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Portugal, United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2433-2021
- AMSORB PLUS PREFILLED G-CAN 1.0L; Carbon dioxide absorbent canister; Product codes AMAB3801 and AMAB3801GE. Used in anesthesia.
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.