Pomona, CA, United States
Alere San Diego Inc. DBA Immunalysis Corporation recalls
1 recall recorded since 2018, most recently on 28 November 2017. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class III Medical devices
- Recall started
- 28 November 2017
- Terminated
- 29 January 2021
- Reason
- The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.
- Distribution
- US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2140-2018
- lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml.