Alcor Scientific, Inc. recalls
1 recall recorded since 2018, most recently on 20 July 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 20 July 2022
- Reason
- Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
- Distribution
- US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1751-2022
- iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-00