Houston, TX, United States
Aggredyne, Inc. recalls
1 recall recorded since 2018, most recently on 23 July 2019. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Completed Class II Medical devices
- Recall started
- 23 July 2019
- Reason
- The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the target RPM for the assay is 0 and the motor output required to achieve this RPM is also 0. Therefore, when the assay is run, the software applies 0 Volts to the motor and performs a 0 RPM assay. Because of this, there is no mixing of the blood with agonist and no moving of aggregates happens in front of the optical detectors. The resulting PAI is 0 and the result is displayed as Low.
- Distribution
- International distribution to the country of South Korea.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1639-2022
- AggreGuide A-100 Instrument
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.