Agfa N.V. recalls

7 recalls recorded since 2018, most recently on 29 July 2024. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)2
Terminated or completed5
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
29 July 2024
Reason
Potential for the front lever chain of the DR 800 table to fail.
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2738-2024

Ongoing Class II Medical devices

Recall started
23 February 2024
Reason
Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Argentina, Australia, Austria, Belgium, Chile, China, Colombia, Czech Republic, Finland, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Mexico, New Zealand, Norway, Peru, Poland, Romania, Russian Federation, Saudi
Initiated
Voluntary: Firm initiated
Recall number
Z-1462-2024

Terminated Class II Medical devices

Recall started
6 April 2020
Terminated
3 April 2023
Reason
Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible
Distribution
US Nationwide distribution including in the sates of FL, IL, IN, LA, MA, MO, OH, SC, VA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2202-2020

Terminated Class II Medical devices

Recall started
10 February 2020
Terminated
30 March 2023
Reason
Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.
Distribution
US Nationwide distribution in the states of FL, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1502-2020

Terminated Class II Medical devices

Recall started
27 June 2019
Terminated
3 June 2020
Reason
The exposed area of fluo exams possibly does not match the active area of the detector within specified limits.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-2056-2019

Terminated Class III Medical devices

Recall started
25 January 2019
Terminated
14 June 2019
Reason
Due to an inhomogeneous coating solution, pinholes can become visible in even dark images. Visibility of the pinholes varies as material from this coating run shows pinholes in different frequencies and intensities
Distribution
US nationwide distribution in the state of OH.
Initiated
Voluntary: Firm initiated
Recall number
Z-1034-2019

Terminated Class II Medical devices

Recall started
15 August 2018
Terminated
23 June 2020
Reason
A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems, changing the pulley and cables if any deterioration in the cables or irregular or excessive pulley wear. If the two cables that support the equipment to its roof anchor break, the equipment might fall and cause harm to the patient, or others.
Distribution
Worldwide Distribution - US Nationwide in the states of CA, DC, FL, IL, IN, KS, LA, MA, MN, MO, NV, NY, OH, OR, SC and TN. Foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-3245-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.