Greenville, SC, United States
AGFA Healthcare Corp. recalls
1 recall recorded since 2018, most recently on 18 November 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 18 November 2025
- Reason
- It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
- Distribution
- U.S. Nationwide distribution.
- Initiated
- FDA Mandated
- Recall number
- Z-0923-2026
- Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800