Advanced Bionics, LLC recalls

9 recalls recorded since 2018, most recently on 30 April 2026. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)4
Terminated or completed5
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
30 April 2026
Reason
Due to incorrect shelf-life expiration date.
Distribution
U.S. Nationwide distribution in the states of CA, CO, GA, IL, IN, KS, MA, MD, MI, MO, MT, NC, TN, WA and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2283-2026

Ongoing Class II Medical devices

Recall started
23 March 2026
Reason
The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.
Distribution
Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canad
Initiated
Voluntary: Firm initiated
Recall number
Z-1950-2026

Ongoing Class II Medical devices

Recall started
27 November 2025
Reason
Behind-the-ear sound processer packaging label is different then included product.
Distribution
Unknown
Initiated
Voluntary: Firm initiated
Recall number
Z-1493-2026

Ongoing Class II Medical devices

Recall started
5 March 2025
Reason
Due to incorrect the battery charger being packaged and distributed. Product labeling does not match the intended product as specified.
Distribution
U.S.: AZ, FL, MI, SC, TN and TX. O.U.S.: N/A
Initiated
Voluntary: Firm initiated
Recall number
Z-1813-2025

Terminated Class III Medical devices

Recall started
26 September 2020
Terminated
4 May 2023
Reason
Mislabeling; It has been determined that some batteries were improperly labeled and may contain a different battery life than indicated.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of NY, IL, MA, NE, CA, HI, LA, MS, AL, FL, SC, OH and the countries of Belgium, Germany, Italy Netherlands, Spain, Colombia, Czech Republic, Iran, Kuwait, Panama, Peru, Taiwan, Italy.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0487-2021, Z-0486-2021

Terminated Class II Medical devices

Recall started
17 February 2020
Terminated
26 January 2023
Reason
hearing performance degradation due to body-fluid entering the device.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Brazil, Bangladesh, Switzerland, China, Australia, India, South Korea, Thailand, Ecuador, Mexico, El Salvador, Guatemala, Malaysia, Myanmar, Taiwan, Chile, Costa Rica, Colombia, Panama, Japan, Argentina, Peru, Singapore
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2033-2020, Z-2032-2020

Terminated Class III Medical devices

Recall started
4 April 2019
Terminated
27 June 2022
Reason
The manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the current version of the software, and users are able to continue programming devices with the current software version.
Distribution
US Nationwide distribution in the sates of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI
Initiated
Voluntary: Firm initiated
Recall number
Z-1441-2020

Terminated Class III Medical devices

Recall started
18 March 2019
Terminated
27 April 2021
Reason
Three materials used in the makeup of the implant were mistakenly left off the list of bio-compatible materials in the firm's previously approved submission to the Pharmaceuticals and Medical Devices Agency (PMDA) of Japan.
Distribution
No US distribution OUS: Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-1259-2020

Terminated Class II Medical devices

Recall started
6 February 2019
Terminated
1 February 2024
Reason
The sound processors were loaded with the incorrect firmware.
Distribution
Distribution was made to NC. Foreign distribution was made to Canada, Argentina, Germany, South Africa, Egypt, Kuwait, and Iran. There was no government/military distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1084-2019, Z-1085-2019, Z-1083-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.