Ad-Tech Medical Instrument Corporation recalls

8 recalls recorded since 2018, most recently on 19 September 2024. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)0
Terminated or completed8
Class I (FDA)0

Completed Class II Medical devices

Recall started
19 September 2024
Reason
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Distribution
US Nationwide distribution in the states of CA, OH.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0221-2025, Z-0219-2025, Z-0220-2025

Terminated Class II Medical devices

Recall started
9 August 2021
Terminated
13 January 2023
Reason
Incorrect version of labels were used.
Distribution
Worldwide distribution - US Nationwide distribution in the states of DC, FL, MI, NY, PA and TX and the countries of Spain, France, Italy, Greece, Great Britain and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2495-2021, Z-2494-2021, Z-2496-2021

Terminated Class II Medical devices

Recall started
27 August 2019
Terminated
5 April 2023
Reason
The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Distribution
NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France
Initiated
Voluntary: Firm initiated
Recall number
Z-2544-2019

Terminated Class II Medical devices

Recall started
18 June 2019
Terminated
20 January 2023
Reason
Supplemental information provided with devices indicates that the subdural electrodes, depth electrodes, and anchor bolts are approved for use in the MR environment, for which they are not approved for such use.
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Worldwide distribution to Albania, Argentina, Australia, Austria
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2406-2019, Z-2407-2019, Z-2408-2019

Terminated Class II Medical devices

Recall started
18 June 2019
Terminated
16 February 2021
Reason
On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts.
Distribution
Worldwide distribution - US Nationwide distribution in the state of Georgia and country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2204-2019

Terminated Class II Medical devices

Recall started
16 April 2019
Terminated
27 May 2022
Reason
An Ad-Tech Clinical Specialist, attended a case on April 15, 2019. During the case the surgeon encountered an issue after implanting four Ad-Tech depth electrodes. Upon initial device testing subsequent to implantation it was identified that two of the devices that were labeled as 8 contact depth electrodes (SD08R-SP05X-000) were in fact 6 contact depth electrodes (SD06R-SP05X-000) . There was no impact to the patient due to this incident.
Distribution
Worldwide - Nationwide distribution in the state of Illinois and countries of Russia and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-2396-2019

Terminated Class II Medical devices

Recall started
2 May 2018
Terminated
8 February 2019
Reason
The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Distribution
US Nationwide distribution in the states of CA, FL, IL, MI, NH, NY, OH, OR, PA, SC, UT, WV. Worldwide distribution to Canada, Denmark, Germany, India, Israel, Japan, Russia, South Korea, Taiwan, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2539-2018

Terminated Class II Medical devices

Recall started
20 December 2017
Terminated
12 October 2018
Reason
There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.
Distribution
Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1938-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.