Acist Medical Systems recalls
1 recall recorded since 2018, most recently on 22 January 2021. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)1
Terminated Class I Medical devices
- Recall started
- 22 January 2021
- Terminated
- 30 October 2023
- Reason
- Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence.
- Distribution
- Worldwide - US Nationwide Distribution in the states of AL, AR, CA, CO, FL, IL, KS, KY, LA, MD, MI, MO, NC, NJ, NY, OK, PA, RI, TN; In the countries of India, Italy, Japan, Poland, and United Arab Emirates.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1161-2021
- ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the ACIST HDi System. The ACIST HDi System is intended to be used for ultrasound examination of coronary a
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.