Acclarent, Inc. recalls

3 recalls recorded since 2018, most recently on 21 September 2023. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)2
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
21 September 2023
Reason
When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during surgical procedures of the Ear Nose and Throat (ENT) and ENT skull base surgery, which may cause delayed/prolonged surgery, cerebrospinal fluid leak, visual impairment, or skull base structural damage.
Distribution
US: CA, IL, WI, MA, TX, CO, FL, ID, KS, OH, OR, NE, MO, AZ, HI, TN, NY, ME, SC, NC, CT, DC, MN, LA, OK, PA, AL, IA, MI, SD
Initiated
Voluntary: Firm initiated
Recall number
Z-0127-2024

Ongoing Class II Medical devices

Recall started
12 November 2021
Reason
Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.
Distribution
US: AZ, MI, IA, LA, IL, VA, SC, NY, IN, OR, SD, TX, PA, NE, FL, CT, MN, NV, CA, NJ, UT, AL, MA, WI, TN, WV, OH, NC, KY, HI¿, OK, GA, CO, MO, WY, MS, NH, HI, MD, ID
Initiated
Voluntary: Firm initiated
Recall number
Z-0415-2022

Terminated Class II Medical devices

Recall started
6 March 2019
Reason
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
Distribution
US: WI, IL, NE, CA, OH, NY OUS: None
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0825-2020, Z-0826-2020, Z-0824-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.