Accelerate Diagnostics Inc recalls

3 recalls recorded since 2018, most recently on 17 October 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
17 October 2022
Reason
Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.
Distribution
U.S.: AR, CA, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, MS, MT, NC, ND, NJ, NY, OH, PA, RI, SC, TN, TX, VA, WA, and WV. O.U.S.: Germany, Italy, Kuwait, Latvia, Poland, Portugal, Romania, Saudi Arabia, Spain, and United Arab Emirates
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0345-2023, Z-0346-2023

Terminated Class II Medical devices

Recall started
14 March 2018
Terminated
17 January 2023
Reason
The kits may produce an elevated rate of false positive calls for the Staphylococcus aureus identification probe (SAU).
Distribution
Distributed to IL, TX, CA, SC, and DE.
Initiated
Voluntary: Firm initiated
Recall number
Z-1529-2018

Terminated Class II Medical devices

Recall started
13 March 2017
Terminated
21 May 2021
Reason
Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.
Distribution
Worldwide Distribution - US Nationwide North America and Middle East US: TX, IL, TN, GA, CA, AR, MD, VA, Washington DC, IA, PA, NY, WA, and MO. OUS (Foreign): UK, Spain, Italy, and Germany
Initiated
Voluntary: Firm initiated
Recall number
Z-1065-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.