Abbott Molecular, Inc. recalls

20 recalls recorded since 2018, most recently on 28 April 2025. 14 are listed as ongoing. Status and classification are as published by the FDA.

Recalls20
Ongoing (FDA)14
Terminated or completed6
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
28 April 2025
Reason
There were reports of an increase in reactive negative controls and false positive results.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1749-2025, Z-1748-2025

Ongoing Class II Medical devices

Recall started
20 March 2025
Reason
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1665-2025

Ongoing Class II Medical devices

Recall started
12 September 2024
Reason
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
Distribution
Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA, IN, IL, LA, MA, MI, MN, MO, NC, NE, NJ, NY, OR, PA, SC, TN, TX, VA, WA, and WI; and foreign (OUS) countries of: AUSTRIA, BELGUIM, BULGARIA, CAMBODIA, CANADA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REP., EL SALVADOR
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0233-2025, Z-0234-2025

Ongoing Class II Medical devices

Recall started
20 March 2024
Reason
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
Distribution
Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERM
Initiated
Voluntary: Firm initiated
Recall number
Z-1602-2024

Ongoing Class III Medical devices

Recall started
1 February 2024
Reason
Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.
Distribution
US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1190-2024, Z-1188-2024, Z-1189-2024

Ongoing Class II Medical devices

Recall started
30 October 2023
Reason
Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0407-2024

Ongoing Class II Medical devices

Recall started
9 December 2022
Reason
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
Distribution
x
Initiated
Voluntary: Firm initiated
Recall number
Z-1009-2023

Ongoing Class II Medical devices

Recall started
22 November 2022
Reason
Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0739-2023, Z-0740-2023

Ongoing Class II Medical devices

Recall started
11 November 2022
Reason
The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results.
Distribution
US Distribution to state of: PA, a government account.
Initiated
Voluntary: Firm initiated
Recall number
Z-1349-2023

Ongoing Class II Medical devices

Recall started
12 October 2022
Reason
Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0268-2023

Ongoing Class II Medical devices

Recall started
19 August 2022
Reason
Reports received of an increase of message code 9186 (internal control failed) when using the kit.
Distribution
Distribution was made to AZ, CA, CT, FL, GA, IL, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NV, NY, OK, PA, RI, TN, TX, WA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Chile and Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1679-2022

Completed Class II Medical devices

Recall started
19 July 2022
Reason
Incorrect optical calibration
Distribution
There was only foreign distribution to Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-1518-2022

Ongoing Class II Medical devices

Recall started
4 March 2022
Reason
There is a potential for misquantitation high results for negative samples.
Distribution
US consignees - AZ, CA, IL, IN, MA, MI, MO, NH, NJ, NY, OH, SD, TX, and VA OUS consignees - Australia, Austria, Belgium, Brazil, Canada, Columbia, Czech Republic, El Salvador, Estonia, France, Germany, Israel, Italy, Japan, Latvia, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Por
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0884-2022, Z-0883-2022

Terminated Class II Medical devices

Recall started
11 February 2022
Terminated
14 April 2023
Reason
There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, CT, DC, FL, GA, IL, KY, LA, MA, MI, MO, NC, NJ, NY, OK, RI, SD, TX, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0710-2022

Ongoing Class II Medical devices

Recall started
6 December 2021
Reason
There is a potential for abnormal (non-sigmoidal) amplification curves.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozambiq
Initiated
Voluntary: Firm initiated
Recall number
Z-0460-2022

Ongoing Class II Medical devices

Recall started
6 December 2021
Reason
There is a software issue associated with the ability to properly complete a Field Service Engineer Maintenance and Diagnostics procedure.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozambiq
Initiated
Voluntary: Firm initiated
Recall number
Z-0461-2022

Terminated Class II Medical devices

Recall started
6 December 2021
Terminated
14 August 2026
Reason
There is a software defect that can cause the Amplification Detection clamp CLEAN position to be set too high.
Distribution
Worldwide distribution - US distribution and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozamb
Initiated
Voluntary: Firm initiated
Recall number
Z-0462-2022

Terminated Class II Medical devices

Recall started
6 December 2021
Terminated
30 September 2024
Reason
There is an issue with the installation of updated camera firmware on the system.
Distribution
Distribution in the United States in CA, DC, GA, KY, IL, MD, MI, MN, MS, NH, NJ, NY, RI, SD, TX, WI OUS distribution to Australia, Cambodia, Canada, Columbia, France, Germany, Israel, Italy, Malaysia, Mozambique, Portugal, South Africa, Spain, Switzerland, Taiwan, United Kingdom, and Vietnam
Initiated
Voluntary: Firm initiated
Recall number
Z-0463-2022

Terminated Class I Medical devices

Recall started
2 September 2021
Terminated
23 July 2024
Reason
There is a potential for false positive results.
Distribution
Worldwide Distribution. US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, COLOMBIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MONTENEGRO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVE
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0007-2022, Z-0008-2022, Z-0004-2022, Z-0006-2022, Z-0005-2022

Terminated Class II Medical devices

Recall started
4 August 2021
Terminated
31 May 2023
Reason
Potential for Vysis CLL FISH Probe Kits not detecting 13q deletions in known positive patient samples.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MI, MN, MA, NC, NJ, NM, NY, OH, PA, TN, TX, UT, VA, VT, WA, WI and the countries of Australia, Belgium, Brazil, Canada, China, Costa Rica, Czech Republic, Estonia, Germany, Hong Kong, Israel, Italy, Kaz
Initiated
Voluntary: Firm initiated
Recall number
Z-2576-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.